Flowtron ACS900
Arjo · SCD (Compression)
Open your document
Open ↗
Stuck? Start here
✅ Check the simple stuff first
- Make sure the patient is safe — swap in a backup unit if you need to.
- Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
- Check power, battery, cables, accessories, and settings before ordering parts.
- Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for Flowtron ACS900 (7)
Low pressure / leak source ↗
- Inspect garment and tubing for leaks; replace garment if alarm continues
Kinked/blocked tube source ↗
- Straighten/clear tubing; replace garment if it persists
Garment disconnected source ↗
- Reconnect or replace garment
Garment removed from limb (compliance) source ↗
- Reapply garment, or standby and restart
Battery low source ↗
- Connect to mains; below 10% therapy can't start; <5% pump shuts down
Faulty tubeset source ↗
- Switch off and call service
Service (spanner) icon source ↗
- Service interval expired; therapy can continue, schedule service
🧰 Common checks for SCD (Compression) (2)
Leak / low pressure / kinked tube source ↗
- Inspect garment and tubing; replace garment
Garment disconnected source ↗
- Reconnect or replace garment
📞 Still stuck? Call Arjo