SureTemp Plus
Baxter · Thermometers
Open your document
Open ↗
Welch Allyn SureTemp Plus 690 692 Service Manual
Instructions for use – SureTemp Plus 690 and 692
SureTemp Plus Veterinary Model 690V, User Manual
SureTemp Plus Wall Holder, User Instructions
SureTemp® Plus 690 and 692 Series, Instructions for Use
SureTemp® Plus Series 690 and 692, Instructions for Use
Welch Allyn SureTemp Plus 690 692 Reference Manual
Welch Allyn SureTemp Plus 690 692 User Manual
Welch Allyn SureTemp Plus 690 and 692 Series, Instructions for Use
Welch Allyn SureTemp Plus Series 690/692, User Manual
Welch Allyn Thermometer Manual de servicio SureTemp Plus Series 690 y 692
Welch Allyn Thermometer Service Manual SureTemp Plus 690 and 692 Series
Stuck? Start here
✅ Check the simple stuff first
- Make sure the patient is safe — swap in a backup unit if you need to.
- Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
- Check power, battery, cables, accessories, and settings before ordering parts.
- Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for SureTemp Plus (7)
Battery indications source ↗
- Blinking frame + 'b' arrow = battery out of range; single blinking segment = low
- Replace batteries
'A' with arrow source ↗
- Ambient outside 10–40 °C: move to normal room temp
Probe or instrument with an X source ↗
- Probe X = probe fault, usually replace probe
- Instrument X = internal fault; read code in Biotech mode Error Log
Reading out of tolerance (calibration check) source ↗
- Verify with the 9600 Plus Calibration Tester (01802-110) or the SureTemp Calibration Key (06138-000)
- Field repairs are limited to calibration check, probe/well, battery and unit replacement
Need software version or error log (Biotech mode) source ↗
- 692: place it in the wall holder; hold Recall while pulling the probe, then release
- Recall steps through the categories (device/algorithm version, ID, error log); Mode or C/F changes values; returning the probe exits; times out after 5 min
Battery drains during cleaning or storage source ↗
- Unplug the latching probe connector when removing the well; reconnect until it snaps in
Wrong probe or well, or probe damaged source ↗
- Blue well for oral/axillary (02891-000), red for rectal (02891-100)
- Probe kits: oral 02893-000 (4 ft) / -100 (9 ft); rectal 02892-000 / -100
- Use only Welch Allyn covers (05031-750); biting the tip damages the probe
🔢 SureTemp Plus error codes (15)
| Code | Meaning | Next step |
|---|---|---|
| E0.1 | Excessive heater energy | Retry; then replace probe source ↗ |
| E0.2 | A/D out of range | Keep ambient 10–40 °C; then replace probe source ↗ |
| E0.4 | Probe over temperature (>43.3 °C) | Let tip cool; then replace probe source ↗ |
| E0.5 | No probe temperature rise | Retry; then replace probe source ↗ |
| E0.6 | Probe EEPROM read failure | Fresh batteries; then replace probe source ↗ |
| E0.7 | Probe EEPROM un-initialized | Fresh batteries; then replace probe source ↗ |
| E0.8 | EEPROM CRC failure | Fresh batteries; then contact Welch Allyn source ↗ |
| E0.9 | Bad probe calibration | Use a calibrated probe source ↗ |
| E4.0 | PTB measurement out of range | Retry; then contact Welch Allyn source ↗ |
| E4.1 | RCAL measurement out of range | Retry; then contact Welch Allyn source ↗ |
| E4.2 | Reference supply malfunction | Contact Welch Allyn source ↗ |
| E4.3–E4.7 | Device EEPROM save/handshake/response/transmit/init failure | Fresh batteries; then contact Welch Allyn source ↗ |
| E4.8 | Device not calibrated | Replace with calibrated device source ↗ |
| E5.0 | Heater not working | Retry; then contact Welch Allyn source ↗ |
| E5.2 | Heater fail-safe failure | Contact Welch Allyn source ↗ |
🧰 Common checks for Thermometers (1)
Error code / probe X source ↗
- Fresh batteries, retry, then replace probe
- Check ambient 10–40 °C
📞 Still stuck? Call Baxter
Contact Baxter US
- 📞 Tech support
- 1-800-535-6663, option 2
- 🛠️ Service
- 1-800-445-3720 (Hillrom beds/stretchers/lifts parts, service, tech support)
- [email protected]
- 🌐 Portal
- Support portal ↗
- 🔩 Parts
- Order parts ↗
Welch Allyn monitoring/cardiology: 800-535-6663 opt 2. Hillrom beds service email [email protected], parts [email protected]. Sigma Spectrum pumps: 1-800-843-7867 (catalog.baxter.com/baxterUS/en/contactus). Fee-based tech support for beds/stretchers without contract since 2025-03-31.
Missing a manual, code, or fix for this model? Tell us.