3800 Series Suction Regulators User Manual

Operator's ManualOEM officialP/N IFU-33784EN
Open ↗

Stuck? Start here

✅ Check the simple stuff first
  1. Make sure the patient is safe — swap in a backup unit if you need to.
  2. Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
  3. Check power, battery, cables, accessories, and settings before ordering parts.
  4. Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for 3800 Series (7)

No suction at patient port source ↗

  • Clean or replace clogged supply/patient fittings

Gauge doesn't respond / slow source ↗

  • Check gauge diaphragm seating
  • Clear hole at top of view tube with small probe

Gauge piston discolored source ↗

  • Fluid got inside: disassemble and clean

Won't shut off / high spike source ↗

  • Replace quad ring (cut by dried fluid), then test

High droop or gauge flutter source ↗

  • Clear brass orifice in diaphragm housing

Fails to regulate / erratic gauge source ↗

  • Disassemble and clean piston/stem
  • Seat retaining ring; no end play in view tube

Intermittent timing wrong source ↗

  • Timing screw: CCW faster, CW slower
🧰 Common checks for Suction (1)

No/poor regulation source ↗

  • Most likely aspirated fluid/dirt: clean fittings, disassemble and clean, replace seals
📞 Still stuck? Call Boehringer Laboratories

Contact Boehringer Laboratories US

🛠️ Service
610-860-6498 (customer service / repair RMA)
✉️ Email
[email protected]
🕘 Hours
M-F 7:30am-5pm ET

Main 610-860-6497. Technical issues via web form (choose 'technical').