CADD-Solis
ICU Medical · Infusion Pumps
Open your document
Open ↗
Stuck? Start here
✅ Check the simple stuff first
- Make sure the patient is safe — swap in a backup unit if you need to.
- Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
- Check power, battery, cables, accessories, and settings before ordering parts.
- Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for CADD-Solis (8)
System Fault alarm (red screen with error numbers, two-tone alarm, amber light) source ↗
- Write down the error numbers
- Clear by pressing both soft keys, or a soft key plus the power switch, or by removing the batteries/AC
- If it recurs, return for service; if the protocol library was lost, do not use the pump and reload the library
Blank screen with alarm, or no power after new AAs source ↗
- Batteries were removed or depleted, or the battery door opened while delivering
- Check AA polarity; install 4 new AAs, a charged pack, or connect the AC adapter
Air in-line detected source ↗
- Clamp the line and disconnect from the patient
- Prime the tubing to remove air, then restart
Downstream/upstream occlusion source ↗
- Clear blockage, kink or closed clamp downstream; delivery resumes when cleared
- Upstream: check the reservoir and tubing above the pump
Cassette alarms (not attached properly / partially unlatched / locked but not latched / detached) source ↗
- Close tubing, remove the cassette fully and reattach; verify the cassette type on screen
- Unlock the cassette/keypad lock and reattach
- Replace the cassette if it persists
Check for empty tubing or reservoir source ↗
- Check for an empty container or air under the pump
- If it persists, the pressure sensor is faulty: remove from service and contact ICU Medical
Battery communication lost (CADD-Solis Communication Module) source ↗
- Keep on AC; the infusion continues but the alarm repeats every 2 minutes
- Stop at a safe point and send the pump and module for service
High/standard volume set mismatch source ↗
- Rates above 250 mL/hr need a high-volume set; swap the cassette to match the protocol
🧰 Common checks for Infusion Pumps (5)
Occlusion source ↗
- Check for closed clamps/stopcocks, kinks, IV site patency
- Check pump height vs patient (LVP)
Air-in-line source ↗
- Look for small bubbles; disconnect and re-prime
Door/cassette/syringe loading source ↗
- Close door with set seated; check syringe clamp, flange and plunger holders
Battery/AC source ↗
- Run on AC; replace/maintain battery; battery service alarms → repair
Device/system error source ↗
- Power cycle once; if it repeats remove from service
📞 Still stuck? Call ICU Medical
Contact ICU Medical US
- 📞 Tech support
- 1-800-241-4002, option 3
- [email protected]
Parts: [email protected]. Canada 1-866-488-6088.
Missing a manual, code, or fix for this model? Tell us.