Medfusion 4000

ICU Medical · Infusion Pumps

Open your document

Medfusion ™ Syringe Pump Model 4000 Operating Manual

Operator's ManualOEM officialP/N 10024593-002EN
Open ↗

Stuck? Start here

✅ Check the simple stuff first
  1. Make sure the patient is safe — swap in a backup unit if you need to.
  2. Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
  3. Check power, battery, cables, accessories, and settings before ordering parts.
  4. Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for Medfusion 4000 (8)

Occlusion - Check Infusion Line source ↗

  • Look for closed stopcocks/clamps, kinked tubing, IV site patency
  • If frequent with no real occlusion: tubing may be too small for rate/viscosity; consider FlowSentry or another occlusion setting
  • If it can't be cleared, send for biomed repair

Syringe flange / plunger sensor alarms source ↗

  • Check barrel clamp, flange holder and plunger holders are engaged and move freely
  • If holder misbehaves or alarm persists, remove from service

Motor Not Running / Motor Rate Error / Force Sensor test source ↗

  • Power off and on
  • If it persists, remove from service for repair

'System Advisory - Maintenance is Recommended' source ↗

  • PM reminder set in Biomed menu; send for PM and reset

Syringe loading sensor alarms (Barrel Clamp / Flange / Plunger) source ↗

  • Engage the barrel clamp, flange holder and plunger holders; check they move freely
  • Persistent or sticking holders: remove from service

Frequent occlusion alarms with no real blockage source ↗

  • Set tubing too small for the rate or viscosity; enable FlowSentry or change the occlusion setting in Configuration
  • Use the smallest syringe for the volume (low rates on large syringes delay occlusion detection)

Motor Not Running / Motor Rate Error / Force Sensor Bgnd Test source ↗

  • Power cycle; if it recurs, biomed repair

Biomed Passcode screen appears after a system fault source ↗

  • Remove from service immediately for inspection or repair
🔢 Medfusion 4000 error codes (15)
CodeMeaningNext step
Check Syringe Barrel ClampClamp sensor can't confirm position, or the clamp was resting on the handle at power-upVerify the clamp is on the barrel; Confirm source ↗
Check Syringe Flange SensorSyringe flange not loaded correctlyRe-seat the clamp, flange holder and plunger holders; if the holder sticks, remove from service source ↗
Check Syringe Plunger SensorPlunger not loaded correctlyRe-seat the plunger holders; they must move freely; otherwise service source ↗
Force Sensor Bgnd TestForce sensor background test failedSilence and restart; if it recurs, service source ↗
Invalid Syringe SizeSize not recognized, or the clamp was lifted during deliveryDon't lift the clamp; check the supported sizes; service if it persists source ↗
Motor Not Running / Motor Rate ErrorSensor, mechanical or electrical faultPower cycle; if it persists, service source ↗
Occlusion - Check Infusion LineLine blockedClear stopcocks, clamps and kinks; restart; if frequent with no real occlusion, the set is too small for the rate or viscosity: consider FlowSentry or a different occlusion setting source ↗
Pressure Increasing - Check Infusion LineFlowSentry detects rising pressureClear the line before it reaches the Occlusion alarm source ↗
Restricted Flow - Bolus/Loading Cancelled / Rate ReducedRestricted flow during bolus or loading doseCheck the line for occlusion source ↗
Calculated Rate Out of Range / Invalid Infusion Parameter Combination / Invalid EntryInvalid program or Drug ProgramVerify the entries; have the PharmGuard Toolbox 2 admin fix the library source ↗
Pump is LockedKeypad lockedPress Unlock source ↗
System Advisory: Low Battery / Low Battery ChangeBattery low; alarms again each 1% dropPlug in source ↗
System Advisory: internal battery not workingBattery failed or depletedUse on AC; service soon source ↗
System Advisory: base hardware/software failure, invalid configuration, network settings corruptBase or configuration faultRemove from service as soon as possible source ↗
System Advisory: improper power-down / Time and Date lost / PM dueHousekeepingSchedule service source ↗
🧰 Common checks for Infusion Pumps (5)

Occlusion source ↗

  • Check for closed clamps/stopcocks, kinks, IV site patency
  • Check pump height vs patient (LVP)

Air-in-line source ↗

  • Look for small bubbles; disconnect and re-prime

Door/cassette/syringe loading source ↗

  • Close door with set seated; check syringe clamp, flange and plunger holders

Battery/AC source ↗

  • Run on AC; replace/maintain battery; battery service alarms → repair

Device/system error source ↗

  • Power cycle once; if it repeats remove from service
📞 Still stuck? Call ICU Medical

Contact ICU Medical US

📞 Tech support
1-800-241-4002, option 3
✉️ Email
[email protected]

Parts: [email protected]. Canada 1-866-488-6088.

Source ↗

Missing a manual, code, or fix for this model? Tell us.