Plum 360

ICU Medical · Infusion Pumps

Open your document

Plum 360 ™ Infuser Operating Manual

Operator's ManualOEM officialP/N IFU0000348EN
Open ↗

Plum 360 ™ Infuser Operating Manual

Operator's ManualOEM officialP/N IFU0000343EN
Open ↗

Plum 360 ™ Infuser Operating Manual

Operator's ManualOEM officialP/N IFU0000342EN
Open ↗

Plum 360 ™ Infuser Operating Manual

Operator's ManualOEM officialP/N IFU0000340EN
Open ↗

Plum 360 ™ Infuser Operating Manual

Operator's ManualOEM officialP/N IFU0000140EN
Open ↗

Plum 360 ™ Infuser Operating Manual

Operator's ManualOEM officialP/N 430-98340-004EN
Open ↗

Plum 360 ™ Infuser Operating Manual

Operator's ManualOEM officialP/N 430-98340-002EN
Open ↗

Plum 360™ System Operating Manual

Operator's ManualOEM officialP/N IFU0000140Rev 09/2020EN
Open ↗

Stuck? Start here

✅ Check the simple stuff first
  1. Make sure the patient is safe — swap in a backup unit if you need to.
  2. Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
  3. Check power, battery, cables, accessories, and settings before ordering parts.
  4. Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for Plum 360 (7)

Distal occlusion alarms source ↗

  • Check the distal line for kinks/closed clamps
  • Resolve, then backprime or open/close cassette door
  • If 'pump too high above patient', lower pump nearer heart level

Cassette test failure source ↗

  • Clear proximal occlusion/air
  • Open and close cassette door, backprime
  • Replace administration set

Battery alarms source ↗

  • Plug into AC
  • 'Service battery / replace pump' (N56/N57): send infuser for repair

Proximal occlusion at startup (N183/N184/N185) or during delivery (N188–N191) source ↗

  • Look for kinks and closed clamps; if all clamps are open, air in the cassette can cause backpressure: backprime
  • Open the cassette door completely and close it to re-run the integrity test
  • Secondary syringe needs free space for backprimed fluid; check the syringe size

Proximal air (N230/N231/N232) source ↗

  • Backprime (System Operating Manual section 4-21); check the syringe size against the set guidelines

Distal occlusion auto-restart (N192) source ↗

  • Have the patient move their arm; the alarm clears within 60 s before max retries
  • Otherwise check the distal line for kinks

Lockout (N255) source ↗

  • Enter the code to unlock
🔢 Plum 360 error codes (22)
CodeMeaningNext step
E301Audio alarm failurePower off; replace/repair infuser source ↗
E302Backlight voltage out of rangePower off; replace/repair infuser source ↗
E325Battery voltage above limitPower off and on; replace pump if it continues source ↗
N180Distal occlusion during backprime/cassette checkCheck distal line, resolve, backprime or reopen door source ↗
N186Distal occlusion after max auto-restartsCheck line and site, restart source ↗
N187Too much backpressure (pump too high)Lower pump toward heart level, press START source ↗
N233/N234Distal air-in-line exceeded thresholdOpen cassette door and follow distal air-in-line procedure source ↗
N250Door opened during deliveryClose door with cassette inserted source ↗
N251Cassette test failureResolve occlusion, reopen door, backprime; replace set source ↗
N252Depleted batteryPlug into AC now source ↗
N58Low batteryPlug into AC source ↗
N56/N57Battery/charging circuit needs serviceKeep on AC; send for repair ASAP source ↗
N108Door opened during Delayed StartClose the door with the cassette inserted source ↗
N192Distal occlusion, auto-restart in progressCheck the distal line; it clears itself if resolved within 60 s before max retries source ↗
N102No input for 2 min after power-onAUDIO PAUSED; continue programming source ↗
N190/N191Proximal occlusion line A during deliveryOpen clamps; if open, air backpressure: backprime source ↗
N188/N189Proximal occlusion line B during deliveryCorrect and restart line B, or clear the program or open the door source ↗
N183Prox occlusion line B at startupSyringe on the secondary port with free space; line unclamped; backprime or open/close the door source ↗
N184/N185Prox occlusion line A at startupBackprime or open/close the door fully; check syringe size source ↗
N230/N231/N232Proximal air line A+B / B / A over thresholdBackprime; check syringe size source ↗
N107Bolus complete on line A (Nurse Callback)AUDIO PAUSED source ↗
N255STOP pressed or door opened while lockedEnter the unlock code source ↗
🧰 Common checks for Infusion Pumps (5)

Occlusion source ↗

  • Check for closed clamps/stopcocks, kinks, IV site patency
  • Check pump height vs patient (LVP)

Air-in-line source ↗

  • Look for small bubbles; disconnect and re-prime

Door/cassette/syringe loading source ↗

  • Close door with set seated; check syringe clamp, flange and plunger holders

Battery/AC source ↗

  • Run on AC; replace/maintain battery; battery service alarms → repair

Device/system error source ↗

  • Power cycle once; if it repeats remove from service
📞 Still stuck? Call ICU Medical

Contact ICU Medical US

📞 Tech support
1-800-241-4002, option 3
✉️ Email
[email protected]

Parts: [email protected]. Canada 1-866-488-6088.

Source ↗