ePM Series

Mindray · Patient Monitors

Open your document

Mindray ePM 15/12/10 Compact Monitor Service Manual v1.0

Service ManualOEM officialRev 1.0EN
Open ↗

Stuck? Start here

✅ Check the simple stuff first
  1. Make sure the patient is safe — swap in a backup unit if you need to.
  2. Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
  3. Check power, battery, cables, accessories, and settings before ordering parts.
  4. Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for ePM Series (6)

ECG lead-off or noisy trace source ↗

  • Check electrode contact and leadwire-to-adapter connections
  • 'ECG Signal Invalid' = high skin impedance: re-prep skin

SpO2 'No Sensor' / 'Sensor Off' source ↗

  • Check extension cable at module and sensor-to-cable joint
  • Replace sensor; use only specified sensors

NIBP leak / timeout source ↗

  • Use correct cuff size for patient category
  • Inspect cuff, hose and pump for leaks; run NIBP leakage test
  • Timeout limit 120 s adult/ped, 90 s neo

Battery / power alarms source ↗

  • Charge on external power
  • 'Battery Error', 'Power Board Comm Error' or supply voltage alarms: restart, then service

Recorder problems source ↗

  • Out of paper/door open, head hot: let cool
  • Init/comm error: restart then service

Network: No CMS / IP conflict source ↗

  • Check LAN/WLAN cabling and IP settings
🔢 ePM Series error codes (25)
CodeMeaningNext step
ECG Lead Off / ECG XX Lead OffElectrode detached or leadwire off adapter cableCheck electrodes and leadwire connections source ↗
ECG NoisyNoisy ECG signalLook for noise sources near cable/electrodes; reduce patient motion source ↗
ECG Signal InvalidSkin impedance too highRe-prep/reapply electrodes source ↗
SpO2 Sensor OffSensor off patient or moduleCheck connection; replace sensor if persists source ↗
SpO2 No SensorExtension cable or sensor disconnectedCheck SpO2 cable and sensor; replace sensor source ↗
SpO2 Excess LightAmbient light too strongCover sensor or move away from light source ↗
SpO2 No Pulse / Low Signal Quality / Low PerfusionNo/poor pulse signalCheck sensor position, change site, reduce shivering/motion; replace sensor source ↗
SpO2 Sensor Incompatible / Sensor ErrorWrong or failed sensorUse specified sensor; replace source ↗
T XX Sensor OffTemp probe disconnectedReconnect probe source ↗
NIBP Cuff Loose / Cuff or Airway Leak / Airway LeakLeak in cuff, tubing or pumpCorrect cuff size, reconnect tubing; run NIBP leakage test source ↗
NIBP Airway Error / Cuff OverpressureTubing occluded or kinkedCheck airway; contact service if persists source ↗
NIBP TimeoutNo reading in 120 s (adult/ped) or 90 s (neo)Check patient, connections; replace cuff source ↗
NIBP Cuff and Patient MismatchCuff type doesn't match patient categoryVerify patient category/cuff; check tubing not bent source ↗
XX No Sensor / XX Sensor Error (IBP)IBP cable/transducer off or failedCheck connections; replace transducer source ↗
Low Battery / Critically Low BatteryBattery lowConnect external power to charge source ↗
Battery ErrorBattery may have failedContact service source ↗
Power Board Comm ErrorPower board communication faultRestart; contact service if persists source ↗
XX V Too High / Too Low2.5/3.3/5/12 V supply out of rangeRestart monitor; service source ↗
RT Clock Need Reset / Not ExistReal-time clock faultContact service source ↗
Recorder Init / Comm ErrorRecorder faultRestart; contact service source ↗
Recorder Head HotPrinting too longPause until print head cools source ↗
Recorder out of PaperNo paper or door openLoad paper, close door source ↗
No CMSDisconnected from central stationCheck network connection source ↗
LAN1/WLAN IP Address Conflict / Fail To Get IPNetwork address problemCheck network settings source ↗
Storage Card ErrorStorage card failed or files damagedRestart to format card; contact service source ↗
🧰 Common checks for Patient Monitors (1)

Leads off / no signal (telemetry) source ↗

  • Re-attach ECG leads
  • Confirm transmitter is in coverage and batteries are good