BSM-2500
Nihon Kohden · Patient Monitors
Open your document
Open ↗
Stuck? Start here
✅ Check the simple stuff first
- Make sure the patient is safe — swap in a backup unit if you need to.
- Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
- Check power, battery, cables, accessories, and settings before ordering parts.
- Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for BSM-2500 (3)
NIBP reading wrong or failing source ↗
- Cuff fit: wrong size, too tight or loose, over thick clothing, or deteriorated
- Reading affected by ESU use, movement, vibration, small pulse, low or rapidly changing BP, CPR, slow pulse
- Use only specified cuffs
CO2 Sensor Error / CO2 Change Adapter source ↗
- Check the CO2 sensor kit and replace if needed; no CO2 monitoring while it's shown
Unexpected alarm behaviour across monitors source ↗
- Alarm defaults may differ per monitor or area: verify the alarm master settings and volume
🧰 Common checks for Patient Monitors (1)
Leads off / no signal (telemetry) source ↗
- Re-attach ECG leads
- Confirm transmitter is in coverage and batteries are good