Efficia CM Series

Philips Healthcare · Patient Monitors

Open your document

Philips Efficia CM Series IFU Addendum (SW 4.3+)

Operator's ManualOEM officialEN
Open ↗

Philips Efficia CM Series Patient Monitors Instructions for Use (Masimo, Rel. 4.0)

Operator's ManualOEM officialEN
Open ↗

Stuck? Start here

✅ Check the simple stuff first
  1. Make sure the patient is safe — swap in a backup unit if you need to.
  2. Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
  3. Check power, battery, cables, accessories, and settings before ordering parts.
  4. Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for Efficia CM Series (4)

Masimo SpO2 no values or searching (Masimo-equipped) source ↗

  • Wait 30 s; reconnect; replace or relocate the sensor

Low SIQ or PI messages source ↗

  • Reposition the sensor; choose a better-perfused site

RRa interference source ↗

  • Reduce ambient or high-intensity light; verify the line frequency setting

SpO2/SpCO values doubtful source ↗

  • Raised COHb raises SpO2; MetHb pushes it toward the low-mid 80s: confirm with lab CO-oximetry
🔢 Efficia CM Series error codes (7)
CodeMeaningNext step
PulseOx Sensor InitializingSensor self-checkIf no values within 30 s, reconnect, then replace the sensor source ↗
PulseOx SearchingNo pulse foundAfter 30 s, reconnect; move to a better-perfused site source ↗
PulseOx Demo ModeDemo sensor connectedFit a patient sensor source ↗
Low SpO2 / Pulse / SpHb / SpCO / SpOC SIQLow signal qualityCheck sensor placement source ↗
Low PI Confidence; Low SpHb/SpCO/SpOC PILow perfusionMove to a better-perfused site source ↗
Low RRa SIQLow RRa signalCheck sensor placement source ↗
RRa Interference DetectedStrong ambient light or wrong line-frequency settingShield the sensor; set the correct line frequency source ↗
🧰 Common checks for Patient Monitors (1)

Leads off / no signal (telemetry) source ↗

  • Re-attach ECG leads
  • Confirm transmitter is in coverage and batteries are good
📞 Still stuck? Call Philips Healthcare

Contact Philips Healthcare US

📞 Tech support
1-800-722-9377
🛠️ Service
1-800-722-9377, option 4 then 3 (parts ordering)

Professional Healthcare Product Support incl. on-site maintenance and parts. Canada 1-800-567-1080.

Source ↗

Missing a manual, code, or fix for this model? Tell us.