Stuck? Start here
✅ Check the simple stuff first
- Make sure the patient is safe — swap in a backup unit if you need to.
- Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
- Check power, battery, cables, accessories, and settings before ordering parts.
- Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for BeneVision N17/N15/N12 (4)
SMR can't identify parameter modules source ↗
- Check SMR-to-monitor cable and contact screws
- Swap in known-good module
- Mismatched software: upgrade module/SMR
- Measure 12 V at SMR interface output and contact screws; abnormal → internal rack COM / main board, or SMR comm module
Internal module rack: module doesn't respond source ↗
- Swap known-good module
- Check internal rack COM board → main control board cable
- Upgrade module/COM board software or COM board logic version
- No 12 V → COM board or main board
Alarm LED off or stuck on source ↗
- Check alarm LED board → main control board cable
- Replace alarm LED board, then main board
No audible alarm source ↗
- Main Menu → System → Maintenance (password): check alarm not muted / minimum volume
🧰 Common checks for Patient Monitors (1)
Leads off / no signal (telemetry) source ↗
- Re-attach ECG leads
- Confirm transmitter is in coverage and batteries are good