BeneVision N22/N19
Mindray · Patient Monitors
Open your document
Open ↗
Stuck? Start here
✅ Check the simple stuff first
- Make sure the patient is safe — swap in a backup unit if you need to.
- Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
- Check power, battery, cables, accessories, and settings before ordering parts.
- Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for BeneVision N22/N19 (9)
Won't power on source ↗
- Confirm AC/battery
- Power-supply protection: remove AC and batteries, wait >15 s, restart
- Check power switch & LED board → front housing interface board → main-control coaxial cable, and power module → DCDC board cables (handle coax carefully)
- Replace power switch & LED board, power module, then motherboard
Blank main screen while running source ↗
- Check same interface/coax cables
- Replace LCD
Secondary display dead source ↗
- Check DP1 connector/cable
- If DP cable was hot-plugged, restart main unit
- Hold secondary screen power button 5 s
Touchscreen not responding / inaccurate source ↗
- Hold Main Menu QuickKey >3 s if touch is disabled
- Check touchscreen → touch control board → front housing interface board → main board cables
- Recalibrate touchscreen
SMR rack: power LED on but modules not recognized source ↗
- Restart (module rack was hot-plugged)
- Check SMR cable and contact screws
- Swap module to test; then rack interface board, 8-slot comm board, DCDC board
No audible alarm though alarm LED works source ↗
- Main Menu → System → Maintenance (password) → Alarm: set Minimum Alarm Volume; then Alarm → Setup volume
- Check speaker cable; replace speaker/main board
No nurse call / analog output source ↗
- Main Menu → Analog Output Setup → Analog Output On
- Check DCDC interface board cable
USB devices not working source ↗
- Restart (USB enumeration)
- Check USB Hub board cable; replace hub board
Battery won't charge source ↗
- Swap in known-good battery and charge fully
- Check battery interface board → power module cable; DCDC interface board
🧰 Common checks for Patient Monitors (1)
Leads off / no signal (telemetry) source ↗
- Re-attach ECG leads
- Confirm transmitter is in coverage and batteries are good
📞 Still stuck? Call Mindray
Contact Mindray US
- 📞 Tech support
- 1-877-913-9663
- 🌐 Portal
- Support portal ↗
Outside North America 650-316-3199.
Missing a manual, code, or fix for this model? Tell us.