Infinity

Moog Medical · Feeding Pumps

Open your document

Infinity Operator's Manual - Manuel de l'opérateur (English & French)

Operator's ManualOEM officialMulti-language
Open ↗

Infinity Troubleshooting Guide (English)

Other DocumentOEM officialEN
Open ↗

Stuck? Start here

✅ Check the simple stuff first
  1. Make sure the patient is safe — swap in a backup unit if you need to.
  2. Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
  3. Check power, battery, cables, accessories, and settings before ordering parts.
  4. Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for Infinity (5)

NO FLOW IN source ↗

  • Check tubing for kinks
  • Clean cassette, manually prime to clear
  • Clean pressure-sensor receptacles; check door/latch not cracked

NO FLOW OUT source ↗

  • Check feeding tube for obstruction; don't use PRIME to clear

NO FOOD (air in tubing) source ↗

  • Check formula level; let foamy formula rest 10–15 min
  • Re-prime (disconnected from patient)

LOAD SET source ↗

  • Power off, clean cassette and receptacles, reload set, close door fully

Charging issues source ↗

  • Check wall plug and adapter; green light on charger
  • Test outlet; inspect pins on adapter and pump
🔢 Infinity error codes (1)
CodeMeaningNext step
ER01–ER99Self-test failurePower off/on, verify settings, restart; call Moog if repeated source ↗
🧰 Common checks for Feeding Pumps (1)

No flow / air / load set source ↗

  • Check kinks, clean cassette and sensor area, re-prime, reload set, close door
📞 Still stuck? Call Moog Medical

Contact Moog Medical US

📞 Tech support
1-800-970-2337

CURLIN pump support.

Source ↗

Missing a manual, code, or fix for this model? Tell us.