CURLIN 8000
Moog Medical · Infusion Pumps
Open your document
Open ↗
Stuck? Start here
✅ Check the simple stuff first
- Make sure the patient is safe — swap in a backup unit if you need to.
- Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
- Check power, battery, cables, accessories, and settings before ordering parts.
- Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for CURLIN 8000 (8)
AIR IN LINE DETECTED source ↗
- Silence; disconnect from the patient; PRIME → CONFIRM; hold PRIME (re-press after 5 mL)
- Check that all air is gone between the pump and the patient connector, even if fluid is flowing
- Air limit: OPTIONS/CLINICAL SETTINGS
AIR MAY BE PRESENT – CHECK IV BAG source ↗
- Bag empty or upstream tubing occluded; hang a new bag; prime with the patient disconnected
CHECK TUBING – ADMINISTRATION SET ISSUE (tubing flattened) source ↗
- Remove and reinsert the tubing; open clamps; change the set if it repeats
DOWN STREAM / UP STREAM OCCLUSION source ↗
- Straighten the patient's arm; trace the tubing to the site; open clamps
- Upstream: check the bag-to-pump segment for blockage
Blank screen with alarm and red light / EMPTY BATTERY source ↗
- Power lost: press Power, replace the batteries, RESUME (edits are discarded)
BATTERY ERROR – UNUSABLE BATTERY / BACKUP BATTERY FAILED TO CHARGE source ↗
- Replace the rechargeable pack or use C cells or AC
- Backup battery: power cycle; if it repeats after 2–3 tries, replace the pump
PUMP IS TOO HOT / DOOR IS OPEN source ↗
- Cool the pump; close the door fully before running
Malfunction error code screen source ↗
- Power off with the On/Off button (don't pull power first)
- Power off/on; if it passes startup (both splash screens, 3 lights flash), review the program and resume. If it recurs, record the code and replace the pump.
🔢 CURLIN 8000 error codes (14)
| Code | Meaning | Next step |
|---|---|---|
| 1, 702, 1000, 1251–1302, 1602 | Software failure | Power off/on; if it passes startup (both splash screens, 3 lights flash), review the program and resume. If it recurs, record the code and replace the pump. source ↗ |
| 2, 3, 501, 1202, 1401 | Software update required | Power off/on; if it passes startup (both splash screens, 3 lights flash), review the program and resume. If it recurs, record the code and replace the pump. source ↗ |
| 15 | Configuration update needed | Connect to RxManager Enterprise Solution software source ↗ |
| 16, 1402, 1603 | Error in current patient files | Power off/on; if it passes startup (both splash screens, 3 lights flash), review the program and resume. If it recurs, record the code and replace the pump. source ↗ |
| 52 / 101 / 151 | Red / green / yellow indicator light failure | Power off/on; if it passes startup (both splash screens, 3 lights flash), review the program and resume. If it recurs, record the code and replace the pump. source ↗ |
| 201 | Screen failure | Power off/on; if it passes startup (both splash screens, 3 lights flash), review the program and resume. If it recurs, record the code and replace the pump. source ↗ |
| 251 / 252 | Piezo / main audio failure | Power off/on; if it passes startup (both splash screens, 3 lights flash), review the program and resume. If it recurs, record the code and replace the pump. source ↗ |
| 301 | Keypad failure | Power off/on; if it passes startup (both splash screens, 3 lights flash), review the program and resume. If it recurs, record the code and replace the pump. source ↗ |
| 452, 1101–1201, 1303–1354, 1450–1454, 1551, 1601, 1701 | Hardware failure | Power off/on; if it passes startup (both splash screens, 3 lights flash), review the program and resume. If it recurs, record the code and replace the pump. source ↗ |
| 505 | Flash failure | Power off/on; if it passes startup (both splash screens, 3 lights flash), review the program and resume. If it recurs, record the code and replace the pump. source ↗ |
| 601 | Health check failure | Power off/on; if it passes startup (both splash screens, 3 lights flash), review the program and resume. If it recurs, record the code and replace the pump. source ↗ |
| 602, 706 | System cross-check failure | Power off/on; if it passes startup (both splash screens, 3 lights flash), review the program and resume. If it recurs, record the code and replace the pump. source ↗ |
| 707 | Watchdog failure | Power off/on; if it passes startup (both splash screens, 3 lights flash), review the program and resume. If it recurs, record the code and replace the pump. source ↗ |
| 1501 | Backup battery failure | Power off/on; if it passes startup (both splash screens, 3 lights flash), review the program and resume. If it recurs, record the code and replace the pump. source ↗ |
🧰 Common checks for Infusion Pumps (5)
Occlusion source ↗
- Check for closed clamps/stopcocks, kinks, IV site patency
- Check pump height vs patient (LVP)
Air-in-line source ↗
- Look for small bubbles; disconnect and re-prime
Door/cassette/syringe loading source ↗
- Close door with set seated; check syringe clamp, flange and plunger holders
Battery/AC source ↗
- Run on AC; replace/maintain battery; battery service alarms → repair
Device/system error source ↗
- Power cycle once; if it repeats remove from service
📞 Still stuck? Call Moog Medical