CURLIN 6000 Clinician Instructions (English)

Operator's ManualOEM officialEN
Open ↗

CURLIN 6000 User Manual (English)

Operator's ManualOEM officialEN
Open ↗

Stuck? Start here

✅ Check the simple stuff first
  1. Make sure the patient is safe — swap in a backup unit if you need to.
  2. Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
  3. Check power, battery, cables, accessories, and settings before ordering parts.
  4. Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for CURLIN 6000 (7)

AIR IN LINE source ↗

  • Check AIR SENS setting (2 mL with no filter)
  • Remove air from set/bag; disconnect from patient before priming

DOWN OCCLUSION source ↗

  • Check set from pump to patient: kinks, blocked filter, access patency
  • Consider down-occlusion setting Low→High
  • If door opened to clear: PAUSE → RESUME → RUN

UP OCCLUSION / HIGH UP PRESSURE source ↗

  • Check set between bag and pump; open clamps
  • High up pressure: relieve pressure on the IV bag

REPLACE SET 3 / REPLACE SET 4 / SET NOT INSTALLED source ↗

  • Replace and prime set
  • Set 4: open all clamps, relieve pressure, open/close door
  • Reinstall set correctly

EMPTY BATTERY / LOW MOTOR SUPPLY / MOTOR STALLED source ↗

  • Replace both C-cells or connect AC adapter; recharge battery pack
  • Solid tone + blank screen + red light = batteries dislodged/dead

SYSTEM TIMEOUT source ↗

  • Empty internal lithium battery: send to Moog/licensed service

Any numeric error code source ↗

  • Power off/on; if pump passes start-up tests resume
  • If it recurs, replace pump and report the code
🔢 CURLIN 6000 error codes (25)
CodeMeaningNext step
11Keyboard buffer fullPower off/on; report code to factory source ↗
12Air-in-line test circuit failurePower off/on; report source ↗
13Bad downstream strain beam (pressure sensor DN)Check admin set; power cycle source ↗
14Bad upstream strain beam (pressure sensor UP)Check/replace set and resume source ↗
15Bad encoderPower cycle; report source ↗
16Bad A/D conversionPower cycle; report source ↗
17Bad clockTurn off; call provider source ↗
18Bad air-in-line comparator reading (FPGA)Report to factory source ↗
19Bad door sensorTurn off; report source ↗
20Empty lithium batteryReport; service source ↗
21–24Bad tristate pin / test pin / reference voltage / motor supplyTurn off; report source ↗
25Stalled pump / broken motor shaftPower cycle and resume; report if recurring source ↗
26–27Motor inhibit / supply enable faultTurn off; report source ↗
29Low auxiliary (alarm) batteryConnect AC battery eliminator and power cycle (may need a few times) source ↗
30Bad auxiliary batteryTurn off; report source ↗
31–32Main speaker audio low/highTurn off; report (service: check D31, replace PCB) source ↗
33Auxiliary speaker audio lowConnect AC eliminator, power cycle source ↗
34Auxiliary speaker highReport source ↗
35History log erasedPower cycle; report source ↗
36/39Bad EEPROM calibration data CRCTurn off; report source ↗
37A/D converter did not runReport source ↗
38Air-in-line circuit not testedReport source ↗
41Bad calibration range (EEPROM limit)Report source ↗
45/46Motor control lossPower cycle; report source ↗
76Battery task timeoutPower cycle; report source ↗
🧰 Common checks for Infusion Pumps (5)

Occlusion source ↗

  • Check for closed clamps/stopcocks, kinks, IV site patency
  • Check pump height vs patient (LVP)

Air-in-line source ↗

  • Look for small bubbles; disconnect and re-prime

Door/cassette/syringe loading source ↗

  • Close door with set seated; check syringe clamp, flange and plunger holders

Battery/AC source ↗

  • Run on AC; replace/maintain battery; battery service alarms → repair

Device/system error source ↗

  • Power cycle once; if it repeats remove from service
📞 Still stuck? Call Moog Medical

Contact Moog Medical US

📞 Tech support
1-800-970-2337

CURLIN pump support.

Source ↗