CURLIN 6000
Moog Medical · Infusion Pumps
Open your document
Open ↗
CURLIN 6000 User Manual (English)
Stuck? Start here
✅ Check the simple stuff first
- Make sure the patient is safe — swap in a backup unit if you need to.
- Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
- Check power, battery, cables, accessories, and settings before ordering parts.
- Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for CURLIN 6000 (7)
AIR IN LINE source ↗
- Check AIR SENS setting (2 mL with no filter)
- Remove air from set/bag; disconnect from patient before priming
DOWN OCCLUSION source ↗
- Check set from pump to patient: kinks, blocked filter, access patency
- Consider down-occlusion setting Low→High
- If door opened to clear: PAUSE → RESUME → RUN
UP OCCLUSION / HIGH UP PRESSURE source ↗
- Check set between bag and pump; open clamps
- High up pressure: relieve pressure on the IV bag
REPLACE SET 3 / REPLACE SET 4 / SET NOT INSTALLED source ↗
- Replace and prime set
- Set 4: open all clamps, relieve pressure, open/close door
- Reinstall set correctly
EMPTY BATTERY / LOW MOTOR SUPPLY / MOTOR STALLED source ↗
- Replace both C-cells or connect AC adapter; recharge battery pack
- Solid tone + blank screen + red light = batteries dislodged/dead
SYSTEM TIMEOUT source ↗
- Empty internal lithium battery: send to Moog/licensed service
Any numeric error code source ↗
- Power off/on; if pump passes start-up tests resume
- If it recurs, replace pump and report the code
🔢 CURLIN 6000 error codes (25)
| Code | Meaning | Next step |
|---|---|---|
| 11 | Keyboard buffer full | Power off/on; report code to factory source ↗ |
| 12 | Air-in-line test circuit failure | Power off/on; report source ↗ |
| 13 | Bad downstream strain beam (pressure sensor DN) | Check admin set; power cycle source ↗ |
| 14 | Bad upstream strain beam (pressure sensor UP) | Check/replace set and resume source ↗ |
| 15 | Bad encoder | Power cycle; report source ↗ |
| 16 | Bad A/D conversion | Power cycle; report source ↗ |
| 17 | Bad clock | Turn off; call provider source ↗ |
| 18 | Bad air-in-line comparator reading (FPGA) | Report to factory source ↗ |
| 19 | Bad door sensor | Turn off; report source ↗ |
| 20 | Empty lithium battery | Report; service source ↗ |
| 21–24 | Bad tristate pin / test pin / reference voltage / motor supply | Turn off; report source ↗ |
| 25 | Stalled pump / broken motor shaft | Power cycle and resume; report if recurring source ↗ |
| 26–27 | Motor inhibit / supply enable fault | Turn off; report source ↗ |
| 29 | Low auxiliary (alarm) battery | Connect AC battery eliminator and power cycle (may need a few times) source ↗ |
| 30 | Bad auxiliary battery | Turn off; report source ↗ |
| 31–32 | Main speaker audio low/high | Turn off; report (service: check D31, replace PCB) source ↗ |
| 33 | Auxiliary speaker audio low | Connect AC eliminator, power cycle source ↗ |
| 34 | Auxiliary speaker high | Report source ↗ |
| 35 | History log erased | Power cycle; report source ↗ |
| 36/39 | Bad EEPROM calibration data CRC | Turn off; report source ↗ |
| 37 | A/D converter did not run | Report source ↗ |
| 38 | Air-in-line circuit not tested | Report source ↗ |
| 41 | Bad calibration range (EEPROM limit) | Report source ↗ |
| 45/46 | Motor control loss | Power cycle; report source ↗ |
| 76 | Battery task timeout | Power cycle; report source ↗ |
🧰 Common checks for Infusion Pumps (5)
Occlusion source ↗
- Check for closed clamps/stopcocks, kinks, IV site patency
- Check pump height vs patient (LVP)
Air-in-line source ↗
- Look for small bubbles; disconnect and re-prime
Door/cassette/syringe loading source ↗
- Close door with set seated; check syringe clamp, flange and plunger holders
Battery/AC source ↗
- Run on AC; replace/maintain battery; battery service alarms → repair
Device/system error source ↗
- Power cycle once; if it repeats remove from service
📞 Still stuck? Call Moog Medical
Missing a manual, code, or fix for this model? Tell us.