Hill-Rom Advanta Bed User Manu
Stuck? Start here
✅ Check the simple stuff first
- Make sure the patient is safe — swap in a backup unit if you need to.
- Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
- Check power, battery, cables, accessories, and settings before ordering parts.
- Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for Advanta (7)
Head section won't raise after CPR release / manual lowering source ↗
- Reset CPR: unplug bed, turn screw assembly clockwise with P2217 IV rod at head end until it clicks
- Remove rod, plug in, hold CPR release until motor stops
A motion doesn't work source ↗
- Check head, knee, Hi-Lo and all-motors lockout controls
PPM won't arm source ↗
- Patient must weigh more than 70 lb (32 kg) and be in bed before activation
- Model A (P1600A) and B (P1600B) arm differently: check which siderail controls you have
PPM false or missed Out of Bed alarms (beds without scale) source ↗
- Zero with an empty bed: PPM off → hold Enable+Volume until the lights stop flashing → keep holding and press Mode 5 s → release; don't touch until the lights go out and it beeps (up to 20 s)
- Beds with a scale: zero the scale instead
PPM alarm won't silence source ↗
- Alarms latch until PPM is turned off at the caregiver siderail, even if the patient returns; then re-arm
Nurse call sent when bed unplugged source ↗
- Expected: with PPM on, unplugging sends a nurse call until power returns
- On power loss nurse call runs at least 4 h, then places an uncancellable call
Brake Not Set indicator flashing source ↗
- Brakes not locked: set the central brake pedal; nudge the bed to confirm
🧰 Common checks for Hospital Beds (1)
Function doesn't move source ↗
- Check lockout controls
- Head stuck after manual CPR lowering: reset CPR per IFU
📞 Still stuck? Call Baxter
Contact Baxter US
- 📞 Tech support
- 1-800-535-6663, option 2
- 🛠️ Service
- 1-800-445-3720 (Hillrom beds/stretchers/lifts parts, service, tech support)
- [email protected]
- 🌐 Portal
- Support portal ↗
- 🔩 Parts
- Order parts ↗
Welch Allyn monitoring/cardiology: 800-535-6663 opt 2. Hillrom beds service email [email protected], parts [email protected]. Sigma Spectrum pumps: 1-800-843-7867 (catalog.baxter.com/baxterUS/en/contactus). Fee-based tech support for beds/stretchers without contract since 2025-03-31.
Missing a manual, code, or fix for this model? Tell us.