Welch Allyn ECG UMDN 11407 CP 150 12-lead resting electrocardiograph Service manual Software v

Operator's ManualArchive copyEN
Open ↗

Stuck? Start here

✅ Check the simple stuff first
  1. Make sure the patient is safe — swap in a backup unit if you need to.
  2. Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
  3. Check power, battery, cables, accessories, and settings before ordering parts.
  4. Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for CP 150 (6)

No power on AC, AC LED off source ↗

  • Check AC source and AC fuses
  • Reseat/replace switch power flex and 60 W PSU harness; replace PSU

No power on AC, AC LED lit source ↗

  • Display or SMCE-LCD flex fault: reseat/replace
  • Check I/O board harness and I/O board

Battery won't charge / few reports per charge source ↗

  • Charge; check connections
  • Fewer than 10 reports on full charge = degraded battery
  • Damaged pins: replace battery case; check smart bus harness

Frozen screen source ↗

  • Hold power ≥6 s until blank, then restart

All leads off / bad waveform on simulator source ↗

  • Replace patient cable
  • Reseat ECG harness; replace ECG board

No IP address with DHCP source ↗

  • Verify network
  • Reseat communication board
🧰 Common checks for ECG (2)

Leads off / bad waveform source ↗

  • Swap patient cable, then harness, then ECG board

Power/battery source ↗

  • Check AC fuses; fewer than 10 reports per charge = replace battery
📞 Still stuck? Call Baxter

Contact Baxter US

📞 Tech support
1-800-535-6663, option 2
🛠️ Service
1-800-445-3720 (Hillrom beds/stretchers/lifts parts, service, tech support)
✉️ Email
[email protected]
🌐 Portal
Support portal ↗
🔩 Parts
Order parts ↗

Welch Allyn monitoring/cardiology: 800-535-6663 opt 2. Hillrom beds service email [email protected], parts [email protected]. Sigma Spectrum pumps: 1-800-843-7867 (catalog.baxter.com/baxterUS/en/contactus). Fee-based tech support for beds/stretchers without contract since 2025-03-31.