Stuck? Start here

✅ Check the simple stuff first
  1. Make sure the patient is safe — swap in a backup unit if you need to.
  2. Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
  3. Check power, battery, cables, accessories, and settings before ordering parts.
  4. Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for IQvitals (8)

NIBP Air Leak / Inflation too quick / Unable to determine NIBP source ↗

  • Check cuff size and the hose for leaks and kinks
  • Values must be within sys 50–260 / dia 20–210

NIBP accuracy suspect (Check Calibration) source ↗

  • Set NIBP mode to Spot → Settings → More → password 1590 → Service mode
  • T in a reference manometer with the cuff on a mandrel; test about 250/100/50 mmHg
  • Pass within ±3 mmHg or 2%, plus the reference tolerance (e.g. 96–104 at 100)

Over-pressure protection check source ↗

  • Inflate to 250 and squeeze the bulb on the pressure side
  • Release must occur at 285–315 mmHg

Masimo SpO2 cable/sensor messages source ↗

  • Replace an expired, defective or unrecognized cable or sensor
  • Check Sensor Connection: reseat

Loss of SpO2/NIBP/LCD Comm. (high priority) source ↗

  • Internal module link failure: remove from use; contact Midmark Technical Service for an RA

Battery absent/faulty, or Critical Battery at 10% source ↗

  • Connect AC; if the battery is faulty, call service

Wireless or scale comm lost source ↗

  • Re-pair the wireless device; check the Fairbanks or 625 table scale link

Annual maintenance source ↗

  • Calibration check every 12 months
🔢 IQvitals error codes (13)
CodeMeaningNext step
NIBP Air LeakCuff or tubing leakCheck cuff and tubing connections source ↗
Unable to determine NIBPNo reading, or pressure out of range (sys 50–260, dia 20–210)Check connections and tubing for kinks source ↗
Inflation too quickCuff/tubing connection issueCheck cuff and tubing source ↗
Low Battery / Critical Battery40%/20% / 10% remainingConnect power source ↗
Battery is absent or faultyBattery missing or badCall service source ↗
Loss of SpO2 / NIBP / LCD Comm.Internal module communication failure (high priority)Service source ↗
Loss of wireless pairing / Loss of wireless Comm.Wireless device or board link lostRe-pair the device source ↗
Loss of Fairbanks scale / Chair Comm.; scale not functionalScale module link failedCheck scale connection source ↗
Unable to load configurationDefaults in useReconfigure or service source ↗
Cannot access patient infoPatient data inaccessibleService source ↗
Max number of patient recordsOldest records will be overwrittenExport or delete records source ↗
Cable/Sensor Life Expired, Incompatible, Unrecognized, Defective (Masimo)SpO2 accessory problemReplace the cable or sensor with a compatible one source ↗
Interference Detected / Sensor Off patientLight or motion interference, or sensor offReapply and shield the sensor source ↗
🧰 Common checks for Vital Signs Monitors (2)

Won't power on source ↗

  • Confirm AC cord/outlet and battery charge
  • Check internal power/battery harnesses
  • Measure power supply output against spec (15 V DC on Welch Allyn Connex and Philips VS3)
  • Charge battery fully; replace if still empty

Touchscreen issues source ↗

  • Check whether touch is disabled/locked
  • Calibrate touchscreen
  • Check touchscreen cables before replacing parts
📞 Still stuck? Call Midmark CorporationMidmark Corporation support ↗