Connex Spot Monitor

Baxter · Vital Signs Monitors

Open your document

Welch Allyn Connex Spot Monitor Service Manual

Service ManualOEM officialP/N 80019225EN
Open ↗

2017 Welch Allyn Connex Spot Monitor, Version 1.X, User Manual

Operator's ManualOEM officialEN
Open ↗

2018 Welch Allyn Connex Spot Monitor, Version 1.X, User Manual

Operator's ManualOEM officialEN
Open ↗

Connex Spot Monitor Service and Maintenance Toolset Quick Reference Guide

Other DocumentOEM officialEN
Open ↗

Connex Spot Monitor, Version 1.X User Manual, English

Operator's ManualOEM officialEN
Open ↗

Instructions For Use Amendment - Welch Allyn® Connex® Spot Monitor

Operator's ManualOEM officialEN
Open ↗

Instructions for use - Welch Allyn Connex Spot Monitor

Operator's ManualOEM officialEN
Open ↗

Special information for US only - Welch Allyn Connex Spot Monitor

Other DocumentOEM officialEN
Open ↗

Wall Mount Kit 7000-GCX for Welch Allyn Connex Spot Monitor, Installation Guide

Other DocumentOEM officialEN
Open ↗

Welch Allyn Connex Spot Monitor - Instructions for Use

Operator's ManualOEM officialEN
Open ↗

Stuck? Start here

✅ Check the simple stuff first
  1. Make sure the patient is safe — swap in a backup unit if you need to.
  2. Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
  3. Check power, battery, cables, accessories, and settings before ordering parts.
  4. Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for Connex Spot Monitor (5)

Won't power up source ↗

  • New battery installed: connect AC to wake it
  • Battery discharged: connect AC to charge
  • Check power supply output is 15 V ±0.45 V DC; replace if not
  • Charge battery 5 h; if icon still empty, replace battery
  • Then power switch, then main board

Touchscreen unresponsive source ↗

  • Calibrate touchscreen / run Host Calibration Sequence in Service Tool
  • Replace LCD/touchscreen, then main board

Powered on but display blank source ↗

  • May be in Sleep mode: touch the screen
  • Run Host Calibration Sequence; replace LCD/touchscreen, then main board

No nurse-call alarm signalling source ↗

  • Check alarm limits aren't set too wide
  • Make sure audio alarms aren't silenced
  • Run Host Calibration Sequence; replace main board

Won't connect to Wi-Fi / Bluetooth source ↗

  • Verify monitor network settings and local network
  • Keep device within range; re-pair Bluetooth
  • Run radio functional check, update software, then replace radio / main board
🔢 Connex Spot Monitor error codes (18)
CodeMeaningNext step
050002NIBP not functional (measurement unavailable)Internal fault: service if persists source ↗
050003Unable to determine NIBP: motion or wrong settingsCheck patient mode/settings, limit movement source ↗
050004Excessive artifactCheck connections, restrict movement source ↗
050005Low inflationCheck inflation settings/patient mode source ↗
050006Kinked tubing or transducer calibration failureCheck hose for kinks; service source ↗
050007Deflation dumped too soonCheck settings/patient mode source ↗
050008Not enough stepsCheck connections, restrict movement source ↗
050009Invalid patient info for modeCheck patient mode source ↗
05000ARe-inflation too lateCheck connections, restrict movement source ↗
05000BToo many re-inflate attemptsCheck inflation settings source ↗
05000CCouldn't bleed below safe venous return pressureCheck tubing for kinks and connections source ↗
05000DAir leak in cuff or tubingCheck cuff and hose connections source ↗
05000FAuto-zero failurePressure unstable: check connections; service source ↗
050105–05020ANIBP module communication/message errorsInternal: service source ↗
050503Factory EEPROM checksum errorService source ↗
050504User EEPROM checksum errorCalibrate NIBP; service if persists source ↗
050505A/D converter POST failureService source ↗
050509Calibration signature zeroCalibrate NIBP module source ↗
🧰 Common checks for Vital Signs Monitors (2)

Won't power on source ↗

  • Confirm AC cord/outlet and battery charge
  • Check internal power/battery harnesses
  • Measure power supply output against spec (15 V DC on Welch Allyn Connex and Philips VS3)
  • Charge battery fully; replace if still empty

Touchscreen issues source ↗

  • Check whether touch is disabled/locked
  • Calibrate touchscreen
  • Check touchscreen cables before replacing parts
📞 Still stuck? Call Baxter

Contact Baxter US

📞 Tech support
1-800-535-6663, option 2
🛠️ Service
1-800-445-3720 (Hillrom beds/stretchers/lifts parts, service, tech support)
✉️ Email
[email protected]
🌐 Portal
Support portal ↗
🔩 Parts
Order parts ↗

Welch Allyn monitoring/cardiology: 800-535-6663 opt 2. Hillrom beds service email [email protected], parts [email protected]. Sigma Spectrum pumps: 1-800-843-7867 (catalog.baxter.com/baxterUS/en/contactus). Fee-based tech support for beds/stretchers without contract since 2025-03-31.