ProBP 3400

Baxter · Vital Signs Monitors

Open your document

Welch Allyn ProBP 3400 Digital Blood Pressure Device, Care & Maintenance Instructions

Service ManualOEM officialEN
Open ↗

Directions for use – Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Operator's ManualOEM officialEN
Open ↗

Instructions for use — Welch Allyn Connex ProBP 3400 digital blood pressure device

Operator's ManualOEM officialEN
Open ↗

Welch Allyn Connex ProBP 3400 Digital Blood Pressure Device, User Manual

Operator's ManualOEM officialEN
Open ↗

Welch Allyn ProBP 3400 Digital Blood Pressure Device, Instructions for Use

Operator's ManualOEM officialEN
Open ↗

Welch Allyn ProBP 3400 Digital Blood Pressure Device, User Manuals

Operator's ManualOEM officialEN
Open ↗

Stuck? Start here

✅ Check the simple stuff first
  1. Make sure the patient is safe — swap in a backup unit if you need to.
  2. Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
  3. Check power, battery, cables, accessories, and settings before ordering parts.
  4. Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for ProBP 3400 (8)

Won't power on source ↗

  • Replace with a fully charged battery
  • Battery contact harness: undamaged, correct polarity, seated at J2 on the main board
  • Flex cable seated at J5; replace the front housing (keypad), then the main board

Battery won't charge / short run time source ↗

  • Check external connections and charge for 6 h; replace battery
  • Check the J2 harness; replace the main board

Display blank or cracked, no backlight source ↗

  • Replace the main board (the display is part of it)

Button unresponsive source ↗

  • Check the flex cable at J5; replace the front housing, then the main board

USB or Bluetooth comms fail source ↗

  • Confirm the unit is on; replace the USB cable; test with the Welch Allyn Service Tool
  • Use the IFU Bluetooth troubleshooting first; then replace the main board

Cuff won't inflate, inaccurate or no BP, persistent NIBP errors source ↗

  • Check external connections; replace the NIBP hose and cuff
  • Manifold to pump, valves and transducer; dump valve cable at J10 (main board); pump cable at J80 (BP board)
  • Leak or clogged filter: replace the manifold; then the pump, BP board, main board

Cuff deflates too slowly source ↗

  • Occlusion blocking the valves: replace the manifold; then the BP board or main board

After any repair source ↗

  • Run the full functional test suite in Service Tool Gold Edition
🔢 ProBP 3400 error codes (2)
CodeMeaningNext step
#00000001System EEPROM errorPower cycle; replace main board source ↗
#00000002ModF errorPower cycle; check J22 (BP board) and J1 (main board); replace BP board, then main board source ↗
🧰 Common checks for Vital Signs Monitors (2)

Won't power on source ↗

  • Confirm AC cord/outlet and battery charge
  • Check internal power/battery harnesses
  • Measure power supply output against spec (15 V DC on Welch Allyn Connex and Philips VS3)
  • Charge battery fully; replace if still empty

Touchscreen issues source ↗

  • Check whether touch is disabled/locked
  • Calibrate touchscreen
  • Check touchscreen cables before replacing parts
📞 Still stuck? Call Baxter

Contact Baxter US

📞 Tech support
1-800-535-6663, option 2
🛠️ Service
1-800-445-3720 (Hillrom beds/stretchers/lifts parts, service, tech support)
✉️ Email
[email protected]
🌐 Portal
Support portal ↗
🔩 Parts
Order parts ↗

Welch Allyn monitoring/cardiology: 800-535-6663 opt 2. Hillrom beds service email [email protected], parts [email protected]. Sigma Spectrum pumps: 1-800-843-7867 (catalog.baxter.com/baxterUS/en/contactus). Fee-based tech support for beds/stretchers without contract since 2025-03-31.

Source ↗

Missing a manual, code, or fix for this model? Tell us.