Stuck? Start here
✅ Check the simple stuff first
- Make sure the patient is safe — swap in a backup unit if you need to.
- Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
- Check power, battery, cables, accessories, and settings before ordering parts.
- Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for ProBP 3400 (8)
Won't power on source ↗
- Replace with a fully charged battery
- Battery contact harness: undamaged, correct polarity, seated at J2 on the main board
- Flex cable seated at J5; replace the front housing (keypad), then the main board
Battery won't charge / short run time source ↗
- Check external connections and charge for 6 h; replace battery
- Check the J2 harness; replace the main board
Display blank or cracked, no backlight source ↗
- Replace the main board (the display is part of it)
Button unresponsive source ↗
- Check the flex cable at J5; replace the front housing, then the main board
USB or Bluetooth comms fail source ↗
- Confirm the unit is on; replace the USB cable; test with the Welch Allyn Service Tool
- Use the IFU Bluetooth troubleshooting first; then replace the main board
Cuff won't inflate, inaccurate or no BP, persistent NIBP errors source ↗
- Check external connections; replace the NIBP hose and cuff
- Manifold to pump, valves and transducer; dump valve cable at J10 (main board); pump cable at J80 (BP board)
- Leak or clogged filter: replace the manifold; then the pump, BP board, main board
Cuff deflates too slowly source ↗
- Occlusion blocking the valves: replace the manifold; then the BP board or main board
After any repair source ↗
- Run the full functional test suite in Service Tool Gold Edition
🔢 ProBP 3400 error codes (2)
| Code | Meaning | Next step |
|---|---|---|
| #00000001 | System EEPROM error | Power cycle; replace main board source ↗ |
| #00000002 | ModF error | Power cycle; check J22 (BP board) and J1 (main board); replace BP board, then main board source ↗ |
🧰 Common checks for Vital Signs Monitors (2)
Won't power on source ↗
- Confirm AC cord/outlet and battery charge
- Check internal power/battery harnesses
- Measure power supply output against spec (15 V DC on Welch Allyn Connex and Philips VS3)
- Charge battery fully; replace if still empty
Touchscreen issues source ↗
- Check whether touch is disabled/locked
- Calibrate touchscreen
- Check touchscreen cables before replacing parts
📞 Still stuck? Call Baxter
Contact Baxter US
- 📞 Tech support
- 1-800-535-6663, option 2
- 🛠️ Service
- 1-800-445-3720 (Hillrom beds/stretchers/lifts parts, service, tech support)
- [email protected]
- 🌐 Portal
- Support portal ↗
- 🔩 Parts
- Order parts ↗
Welch Allyn monitoring/cardiology: 800-535-6663 opt 2. Hillrom beds service email [email protected], parts [email protected]. Sigma Spectrum pumps: 1-800-843-7867 (catalog.baxter.com/baxterUS/en/contactus). Fee-based tech support for beds/stretchers without contract since 2025-03-31.
Missing a manual, code, or fix for this model? Tell us.