MRx
Philips Healthcare · Defibrillators
Open your document
Open ↗
Heartstart MRx Service Manual 2005
HeartStart MRx Reference Card
Heartstart MRx Calibration Instructions
Philips Defibrillator HeartStart MRx M3535A M3536A Instructions for Use
Philips Defibrillator HeartStart MRx M3535A M3536A Instructor Guide
Philips Defibrillator HeartStart MRx M3535A M3536A User Training Workbook
Philips Defibrillator Philips Heartstart MRx Defibrillator Maintenance Guide
Stuck? Start here
✅ Check the simple stuff first
- Make sure the patient is safe — swap in a backup unit if you need to.
- Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
- Check power, battery, cables, accessories, and settings before ordering parts.
- Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for MRx (8)
Won't turn on / Batteries Low / Shutting Down source ↗
- Insert a fully charged battery or connect AC/DC; Batteries Low means less than 10 min of monitoring or six 200 J shocks
- Charge indicators inaccurate: calibrate the battery; short life: replace it
Poor ECG (noise or wandering baseline) or flat line with no Leads Off source ↗
- Fresh, in-date electrodes or pads with skin prep; move away from RFI sources
- Run the Operational Check with the cable, then without it; if it passes without, replace the cable, otherwise service
No Shock Delivered: Replace Pads / Press Pads Firmly / Low-High Impedance source ↗
- Low impedance: replace the pads; high impedance: press pads firmly
- Paddles: improve skin contact; replace the paddles if it persists
Slow charge to selected energy source ↗
- Running on AC/DC with no or a low battery: install a charged battery
Pacing Stopped [Leads Off] / [Pads Off] source ↗
- Reapply the electrodes or pads and check the cables, then press [Resume Pacing]
NBP Cuff Not Deflated / Interrupted / Measurement Failed; cuff won't inflate source ↗
- Remove and deflate the cuff; check cuff size, placement and tubing; replace the cuff
- Calibration is due yearly or every 10,000 cycles; end of life is 50,000 cycles
Printer: paper won't move, faint print, missing dots, grinding source ↗
- Reload the paper or clear the jam; replace wet paper; latch the door
- Clean the printhead; let it cool if hot
- Printer Malfunction: power off for 15 s; otherwise service
Operational Check failures (Charge/Shock Button, Therapy Knob, Shock Equip) source ↗
- Remove from use and call service
- Therapy Cable Failure during the check: attach a therapy cable and rerun
🔢 MRx error codes (15)
| Code | Meaning | Next step |
|---|---|---|
| Device Error. Service Required. | Failure detected at start-up | Take out of use; call service source ↗ |
| All Settings Have Been Set To Factory Default Values | Corrupt or incomplete configuration | Reload the configuration file source ↗ |
| Fan Failure | Internal fan not running | Service source ↗ |
| Internal Memory Failure | Card reformatted after a failed self-test | Ready again; prior data was erased source ↗ |
| Event Storage Full | Over capacity or the 12 h limit | Print or copy the Event Summary and start a new case source ↗ |
| Power Supply Failure INOP | Internal power supply fault | Remove from use source ↗ |
| Replace Clock Battery | Internal clock battery | Service source ↗ |
| Shock Equip Malfunction INOP | Hardware prevents a shock | Remove from use source ↗ |
| Pads ECG Cable Failure / ECG Cable Failure INOP | Lead wire shorted to ground | Replace the cable; Operational Check source ↗ |
| ECG Equip Malfunction INOP | Hardware failure | Operational Check without the cable; service if the Leads ECG test fails source ↗ |
| Pacer Equip Malfunction INOP | Pacing hardware | Service source ↗ |
| NBP Calibration Overdue / NBP Equip Malfunction / NBP Service Required | Calibration due, NBP hardware fault, or 50k cycles | Service the NBP module; don't use NBP source ↗ |
| Maximum Energy 50J | Internal paddles | Select ≤50 J source ↗ |
| R-Wave Not Detected | Sync cardioversion | Pick a lead with a clear QRS source ↗ |
| Printer Malfunction / Check Printer / Incompatible Printer | Printer fault, no paper or door open, or fonts | Power cycle 15 s; load paper; Asian fonts need the 75 mm printer source ↗ |
🧰 Common checks for Defibrillators (1)
Won't charge or deliver shock source ↗
- Isolate controls: try paddles vs. pads vs. front-panel buttons
- Check pad/paddle contact and cables (impedance disarm)
- Check battery / run on AC; slow charge often = battery
- Read stored failure codes in service mode before replacing therapy module
📞 Still stuck? Call Philips Healthcare
Contact Philips Healthcare US
- 📞 Tech support
- 1-800-722-9377
- 🛠️ Service
- 1-800-722-9377, option 4 then 3 (parts ordering)
Professional Healthcare Product Support incl. on-site maintenance and parts. Canada 1-800-567-1080.
Missing a manual, code, or fix for this model? Tell us.