LIFEPAK 20
Stryker · Defibrillators
Open your document
Open ↗
CMM PIP Checklist 3312981-001_B
CMM PIP-TCP Manual 3312981-000_B
LP20e Service BOM 3343496_AG
PIP Checklist LP20/20e 3201896-002_C
PIP Procedure LP20e 3201896-000_C
TCP Procedure LP20e 3201896-001_C
Stuck? Start here
✅ Check the simple stuff first
- Make sure the patient is safe — swap in a backup unit if you need to.
- Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
- Check power, battery, cables, accessories, and settings before ordering parts.
- Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for LIFEPAK 20 (9)
Service indicator LED on source ↗
- Record codes (date/time/extension) from Service Log, select CLEAR LOG, power off
- Run the PIP; if it passes code was likely EMI/ESD
- If LED returns, look up code in Service Code Table; repeat PIP after repair
Beeps when turned off source ↗
- Connect to AC power or disable the AC Loss Alert alarm
Display on but no power-on LED / frozen at power-on / continuous reset source ↗
- Check W17/W18 UI flex cables
- A06 UI PCB, A03 Therapy PCB; continuous reset → A01 System PCB
Printer not printing / light print source ↗
- Run printer test, use proper paper
- Check 3.3 V on pins 14 & 16 of J38 test connector on A06 UI PCB: higher → replace W14 printer flex; lower → A12 printer or A06
No backup battery operation source ↗
- Replace E09 battery; check E08 power module
Auto/user test can't complete; energy out of tolerance source ↗
- Rerun with proper test-load shorting; check test plug/shorting bar and QUIK-COMBO cable continuity
- Check W01 Therapy Connector; perform defibrillator calibration
- Abnormal energy: E06 inductive resistor should be <5 Ω
Low patient impedance on 3:00 AM test source ↗
- If tested into 50 Ω load, rerun with correct test plug; then PIP
- If it fails, A01 System PCB
No sync mark source ↗
- Paddles channel: therapy cable, W01
- Lead channel: noisy ECG, ECG cable, W06 ECG connector, A02 Patient Parameter PCB
Pacer won't turn on source ↗
- Verify manufacturer configuration bit
- Run keypad test; elastomer keypad / A06 UI PCB
🧰 Common checks for Defibrillators (1)
Won't charge or deliver shock source ↗
- Isolate controls: try paddles vs. pads vs. front-panel buttons
- Check pad/paddle contact and cables (impedance disarm)
- Check battery / run on AC; slow charge often = battery
- Read stored failure codes in service mode before replacing therapy module
📞 Still stuck? Call Stryker