LIFEPAK 15

Stryker · Defibrillators

Open your document

Service Manual LP15 V1/V2 3300117_M

Service ManualOEM officialP/N 3300117Rev MEN
Open ↗

Service Manual LP15 V4 3316925_AL

Service ManualOEM officialP/N 3316925Rev ALEN
Open ↗

LP15 V2 Service BOM 3347648_A

Parts ListOEM officialP/N 3347648Rev AEN
Open ↗

LP15 V4 Service BOM 3343060_AK

Parts ListOEM officialP/N 3343060Rev AKEN
Open ↗

LP15 V4 Service Configurations_BA

Other DocumentOEM officialEN
Open ↗

PIP Checklist LP15 3207841-002_AN

PM ProcedureOEM officialP/N 3207841-002Rev ANEN
Open ↗

PIP Procedure LP15 3207841-000_AN

PM ProcedureOEM officialP/N 3207841-000Rev ANEN
Open ↗

TCP Procedure LP15 3207841-001_AN

PM ProcedureOEM officialP/N 3207841-001Rev ANEN
Open ↗

Stuck? Start here

✅ Check the simple stuff first
  1. Make sure the patient is safe — swap in a backup unit if you need to.
  2. Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
  3. Check power, battery, cables, accessories, and settings before ordering parts.
  4. Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for LIFEPAK 15 (9)

No power on source ↗

  • Install fully charged, maintained batteries
  • Inspect battery pins for damage
  • Check W04 System/Interface PCB cable; possible A03 Power PCB

Service LED on source ↗

  • Enter Service mode and read the Service Log; process codes per 'Processing Service Log Codes'
  • LED shows a new code was written — review log even if LED goes off after restart

'MAINTENANCE DUE' message source ↗

  • Reset the maintenance prompt interval in Service mode

Printer: 'CHECK PRINTER', missing dots/characters source ↗

  • Use proper paper and load correctly
  • Clean printhead and paper sensor
  • Check A12 Printer Assembly; 25/12.5 mm marker spacing → TCP Printer Calibration

No energy discharge / 'ABNORMAL ENERGY DELIVERY' source ↗

  • Verify test setup; check QUIK-COMBO therapy cable or hard paddles
  • Check A13 Transfer Relay connection at P22
  • Possible W11 therapy cable, A04 Therapy PCB, A22 Biphasic PCB, A14 inductive resistor, A15 capacitor
  • Energy out of tolerance → TCP Defibrillator Energy Calibration

Sync cardioversion problems source ↗

  • No sync mark: adjust ECG size
  • Sync discharge >60 ms or no transfer: check A10 main keypad, cables, A01 System PCB

No pacer output source ↗

  • Check A13 Transfer Relay/P22
  • Possible A04 Therapy PCB

ECG gain out of tolerance / wrong messages (3/5/12-lead) source ↗

  • Check ECG cable
  • Possible W07 ECG connector cable or A01 System PCB

Display issues / keys source ↗

  • Run Service mode TESTS > Display Pixels Test
  • Possible A11 LCD, A09/A10 keypads, A05 Interface PCB
🔢 LIFEPAK 15 error codes (14)
CodeMeaningNext step
CA 1System communications/processing errorCheck cables/harnesses; clear Service Log and run PIP (possible EMI/ESD); possible A01 System PCB source ↗
CA 2Real-time clock failurePossible NVRAM/coin battery failure on A01 System PCB source ↗
CA 6Therapy processor errorCheck cables; possible A04 Therapy PCB source ↗
CA 7Power processor errorCheck cables; possible A03 Power PCB source ↗
CA 8Manufacturing configuration bit not setFactory service required; possible A01 NVRAM source ↗
CA 10Defibrillator out of calibrationTCP Defibrillator Energy Calibration then PIP; check A04, A22/A14 source ↗
CA 11Configuration or keypad errorCheck config vs installed hardware; A09/A10 keypads source ↗
CA 13Printer errorA12 Printer Assembly; cables source ↗
CA 15Pacer out of calibrationTCP Pacer Self-Calibration then PIP; A04 source ↗
CA 17Biphasic fault / unknown energyCheck cables; A22 Biphasic PCB, A04 source ↗
CA 19Dump error (a02c)Check cables; A04, A22, A01 source ↗
CA 20ECG errorCheck cables; A01 System PCB source ↗
CA 28Recommended software versionContact Stryker field service source ↗
CA 31Data managementClear Data Management memory source ↗
🧰 Common checks for Defibrillators (1)

Won't charge or deliver shock source ↗

  • Isolate controls: try paddles vs. pads vs. front-panel buttons
  • Check pad/paddle contact and cables (impedance disarm)
  • Check battery / run on AC; slow charge often = battery
  • Read stored failure codes in service mode before replacing therapy module
📞 Still stuck? Call Stryker

Contact Stryker US

📞 Tech support
1-800-327-0770, option 2
✉️ Email
[email protected]
🕘 Hours
M-F 8am-7pm ET
🌐 Portal
Support portal ↗

Emergency Care (LIFEPAK): +1 800 327 0770; software/connectivity +1 844 999 2787 (https://www.stryker.com/us/en/emergency-care/contact.html).

Source ↗