SureSigns VS3
Philips Healthcare · Vital Signs Monitors
Open your document
Open ↗
Stuck? Start here
✅ Check the simple stuff first
- Make sure the patient is safe — swap in a backup unit if you need to.
- Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
- Check power, battery, cables, accessories, and settings before ordering parts.
- Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for SureSigns VS3 (9)
Runs on battery but not AC source ↗
- Check power cord plugged in / not broken
- Check fuse
- Check power supply output 15 V; check DC-in cable and AC connector
- Then front panel keypad, then main board
Runs on AC but not battery source ↗
- Charge battery
- Reseat battery and battery cable
- Replace battery connector board, then main board
Won't turn on with either AC or battery source ↗
- Work through the AC-only and battery-only steps; then replace the LCD
Battery service required source ↗
- Max Error over 8%, or Full Capacity at half of Design Capacity: recondition the battery
Screen blank or distorted, or freezes at boot source ↗
- Blank: check Standby (press On/Standby), then the backlight/LCD/main board connections
- Distorted graphics on white: reseat the LCD signal cable; replace the LCD
- Freezes at the opening screen or colour bar: replace the main board
AUDIO FAILED (flashing red/white box), or no audible alarms source ↗
- Check the speaker connection; clear it by running the speaker test or clearing the error log
- Check Audio Pause/Off; replace the speaker, then the main board
NBP cuff won't inflate, or no or unreliable readings source ↗
- Straighten kinked tubing; replace a leaking cuff; check cuff size and initial pressure setting
- Replace the blocked NBP pump filter, then the pump module or main board; recalibrate (NBP Calibration Procedure)
Temperature not displayed or unreliable source ↗
- Probe connected and seated in its well; predictive range is 15–45°C
- Tympanic high: torn or loose cover; low: obstruction
- Temporal low: moist skin, readings under 30 s apart, curved sweep, dirty lens
- Calibrate or test the module; replace the probe, front end board or module
Only one Charging LED colour source ↗
- Replace the membrane switch assembly
🔢 SureSigns VS3 error codes (14)
| Code | Meaning | Next step |
|---|---|---|
| 257 | System error: battery charger power failure | Replace battery; then main board; then AC power module source ↗ |
| 261 | System error: battery overcharged | Run without AC; then replace battery; then main board source ↗ |
| 264 | System error: battery over temperature | Shut down and restart; then replace battery; then main board source ↗ |
| 817 | NBP FPGA FIFO overrun (software) | Informational, no action required source ↗ |
| 818 (2) | NBP power-on self test failed (A/D or transducer offset) | Replace NBP module; then main board source ↗ |
| 818 (15) | NBP unrecoverable firmware error | Replace main board source ↗ |
| 818 (25) | NBP hardware malfunction, no data | Replace NBP module; then main board source ↗ |
| 818 (101) | NBP task overrun | Shut down and restart; if it persists replace main board source ↗ |
| 818 (102) | NBP interrupt stack overwritten | Shut down and restart; if it persists replace main board source ↗ |
| 818 (103) | NBP user stack overwritten | Shut down and restart; if it persists replace main board source ↗ |
| 818 (104) | NBP data validity check failed | Shut down and restart; if it persists replace main board source ↗ |
| 818 (106) | Can't write NBP init flag to NV memory | Shut down and restart; if it persists replace main board source ↗ |
| 818 (107) | Can't verify NBP init flag in NV memory | Shut down and restart; if it persists replace main board source ↗ |
| 818 (109) | Can't write NBP config to NV memory | Shut down and restart; if it persists replace main board source ↗ |
🧰 Common checks for Vital Signs Monitors (2)
Won't power on source ↗
- Confirm AC cord/outlet and battery charge
- Check internal power/battery harnesses
- Measure power supply output against spec (15 V DC on Welch Allyn Connex and Philips VS3)
- Charge battery fully; replace if still empty
Touchscreen issues source ↗
- Check whether touch is disabled/locked
- Calibrate touchscreen
- Check touchscreen cables before replacing parts
📞 Still stuck? Call Philips Healthcare
Contact Philips Healthcare US
- 📞 Tech support
- 1-800-722-9377
- 🛠️ Service
- 1-800-722-9377, option 4 then 3 (parts ordering)
Professional Healthcare Product Support incl. on-site maintenance and parts. Canada 1-800-567-1080.
Missing a manual, code, or fix for this model? Tell us.