LIFEPAK 35
Stryker · Defibrillators
Open your document
Open ↗
LIFEPAK 35 Service Manual 3324158_AF
LIFEPAK 35 BOM 3342155_AN
LIFEPAK 35 PIP Checklist 3340124-002_AF
LIFEPAK 35 Performance Inspection Procedure (PIP) 3340124-000_AH
LIFEPAK 35 Service Configurations 3353944_C
LIFEPAK 35 Test Calibration Procedure (TCP) 3340124-001_AH
Stuck? Start here
✅ Check the simple stuff first
- Make sure the patient is safe — swap in a backup unit if you need to.
- Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
- Check power, battery, cables, accessories, and settings before ordering parts.
- Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for LIFEPAK 35 (8)
No power source ↗
- Use a charged LIFEPAK FLEX battery or AC
- Inspect/clean battery pins; service if bent/loose
- Check power cord and adapter
POWER LED on, screen blank, unresponsive source ↗
- Hold POWER ~5 s until LED off, restart
- If it won't turn off, remove both batteries and AC, then restart
Battery messages source ↗
- CANNOT CHARGE BATTERY: swap battery/cord; else service
- BATTERY X: REMOVE BATTERY FROM SERVICE: end of life, replace
- BATTERY X: INVALID BATTERY: replace; use only LIFEPAK FLEX
- Fuel gauge dark: charge battery (new or deeply depleted)
Turns off unexpectedly source ↗
- Low battery/high draw: connect AC, restart
- Move RF equipment away
TOUCH SCREEN DISABLED – USE SPEED DIAL source ↗
- Use speed dial; power-cycle; service if persists
OVERHEATING – ALLOW DEVICE TO COOL DOWN source ↗
- Remove from sun/coverings, cool; service if persists
Printer messages source ↗
- OUT OF PAPER: new roll
- DOOR OPEN: latch door
- ERROR DETECTED: clear jam/misfeed
Service LED source ↗
- Power-cycle (creates new record); if not cleared remove from use and report
🧰 Common checks for Defibrillators (1)
Won't charge or deliver shock source ↗
- Isolate controls: try paddles vs. pads vs. front-panel buttons
- Check pad/paddle contact and cables (impedance disarm)
- Check battery / run on AC; slow charge often = battery
- Read stored failure codes in service mode before replacing therapy module
📞 Still stuck? Call Stryker
Contact Stryker US
- 📞 Tech support
- 1-800-327-0770, option 2
- [email protected]
- 🕘 Hours
- M-F 8am-7pm ET
- 🌐 Portal
- Support portal ↗
Emergency Care (LIFEPAK): +1 800 327 0770; software/connectivity +1 844 999 2787 (https://www.stryker.com/us/en/emergency-care/contact.html).
Missing a manual, code, or fix for this model? Tell us.