LIFEPAK 35

Stryker · Defibrillators

Open your document

LIFEPAK 35 Service Manual (AU) 3351283_A

Service ManualOEM officialP/N 3351283Rev AEN
Open ↗

LIFEPAK 35 Service Manual 3324158_AF

Service ManualOEM officialP/N 3324158Rev AFEN
Open ↗

LIFEPAK 35 BOM 3342155_AN

Parts ListOEM officialP/N 3342155Rev ANEN
Open ↗

LIFEPAK 35 PIP Checklist 3340124-002_AF

PM ProcedureOEM officialP/N 3340124-002Rev AFEN
Open ↗

LIFEPAK 35 Performance Inspection Procedure (PIP) 3340124-000_AH

PM ProcedureOEM officialP/N 3340124-000Rev AHEN
Open ↗

LIFEPAK 35 Service Configurations 3353944_C

Other DocumentOEM officialP/N 3353944Rev CEN
Open ↗

LIFEPAK 35 Test Calibration Procedure (TCP) 3340124-001_AH

PM ProcedureOEM officialP/N 3340124-001Rev AHEN
Open ↗

Stuck? Start here

✅ Check the simple stuff first
  1. Make sure the patient is safe — swap in a backup unit if you need to.
  2. Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
  3. Check power, battery, cables, accessories, and settings before ordering parts.
  4. Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for LIFEPAK 35 (8)

No power source ↗

  • Use a charged LIFEPAK FLEX battery or AC
  • Inspect/clean battery pins; service if bent/loose
  • Check power cord and adapter

POWER LED on, screen blank, unresponsive source ↗

  • Hold POWER ~5 s until LED off, restart
  • If it won't turn off, remove both batteries and AC, then restart

Battery messages source ↗

  • CANNOT CHARGE BATTERY: swap battery/cord; else service
  • BATTERY X: REMOVE BATTERY FROM SERVICE: end of life, replace
  • BATTERY X: INVALID BATTERY: replace; use only LIFEPAK FLEX
  • Fuel gauge dark: charge battery (new or deeply depleted)

Turns off unexpectedly source ↗

  • Low battery/high draw: connect AC, restart
  • Move RF equipment away

TOUCH SCREEN DISABLED – USE SPEED DIAL source ↗

  • Use speed dial; power-cycle; service if persists

OVERHEATING – ALLOW DEVICE TO COOL DOWN source ↗

  • Remove from sun/coverings, cool; service if persists

Printer messages source ↗

  • OUT OF PAPER: new roll
  • DOOR OPEN: latch door
  • ERROR DETECTED: clear jam/misfeed

Service LED source ↗

  • Power-cycle (creates new record); if not cleared remove from use and report
🧰 Common checks for Defibrillators (1)

Won't charge or deliver shock source ↗

  • Isolate controls: try paddles vs. pads vs. front-panel buttons
  • Check pad/paddle contact and cables (impedance disarm)
  • Check battery / run on AC; slow charge often = battery
  • Read stored failure codes in service mode before replacing therapy module
📞 Still stuck? Call Stryker

Contact Stryker US

📞 Tech support
1-800-327-0770, option 2
✉️ Email
[email protected]
🕘 Hours
M-F 8am-7pm ET
🌐 Portal
Support portal ↗

Emergency Care (LIFEPAK): +1 800 327 0770; software/connectivity +1 844 999 2787 (https://www.stryker.com/us/en/emergency-care/contact.html).

Source ↗

Missing a manual, code, or fix for this model? Tell us.