Infusomat Space
B. Braun · Infusion Pumps
Open your document
Open ↗
B. Braun Infusomat Space Infusion Pump Service Manual
Infusomat Space Large Volume Pump IFU (software G)
Stuck? Start here
✅ Check the simple stuff first
- Make sure the patient is safe — swap in a backup unit if you need to.
- Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
- Check power, battery, cables, accessories, and settings before ordering parts.
- Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for Infusomat Space (7)
Device alarm source ↗
- Infusion stops: switch off and on
- If repeated, clamp, disconnect patient, remove set and send to service
'Pressure high' (occlusion) source ↗
- Check tubing for kinks/damage and IV/filter patency
- Raise occlusion pressure if appropriate
'Air bubble' / 'Accumulated air' source ↗
- Check set for small bubbles; disconnect and re-prime if needed
'Drive blocked' source ↗
- Excess pressure: disconnect patient and reinsert the line
Battery problems source ↗
- 'Battery empty': connect AC and/or swap pack
- 'Battery cover removed': push cover until it clicks
- 'No battery inserted': pump can't run without a pack
- Run monthly battery maintenance (~12 h); blinking battery >1 h on AC = have it checked
'Calibrate device' or wrench icon with yellow/red/blue LEDs source ↗
- Recalibrate via service program
- Wrench = service mode: not for patient use
'Communication error' pre-alarm source ↗
- Incompatible/defective device in the stack: have system checked
🧰 Common checks for Infusion Pumps (5)
Occlusion source ↗
- Check for closed clamps/stopcocks, kinks, IV site patency
- Check pump height vs patient (LVP)
Air-in-line source ↗
- Look for small bubbles; disconnect and re-prime
Door/cassette/syringe loading source ↗
- Close door with set seated; check syringe clamp, flange and plunger holders
Battery/AC source ↗
- Run on AC; replace/maintain battery; battery service alarms → repair
Device/system error source ↗
- Power cycle once; if it repeats remove from service
📞 Still stuck? Call B. Braun
Contact B. Braun US
- 🛠️ Service
- 1-800-227-2862 (general customer support)
- [email protected]
Page lists AIS tech support 972-466-5068 and Renal tech support 1-800-621-0445; no dedicated infusion-pump tech line on that page.
Missing a manual, code, or fix for this model? Tell us.