ELI 250
Baxter · ECG
Open your document
Open ↗
Microsoft Word - 9516-163-50-ENG_REV_D ELI 250-210 SM
Mortara ELI 250 ELI 210 Service Manual
Mortara ELI 250 User Manual
Stuck? Start here
✅ Check the simple stuff first
- Make sure the patient is safe — swap in a backup unit if you need to.
- Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
- Check power, battery, cables, accessories, and settings before ordering parts.
- Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for ELI 250 (8)
Backlight on with AC but nothing displayed source ↗
- Check the AC fuses
- Does it boot on battery? Yes: AC supply fault, replace motherboard. No: replace motherboard and check battery
- Reseat, then replace, the keyboard/LCD cable
LCD blank but key tones sound source ↗
- LCD fault: replace the LCD
Works on AC but not on battery source ↗
- Check the battery fuse and connections; replace battery
- Power Supply Test: off current <40 µA, on current <250 mA, charger 13.0–14.0 VDC with the AC LED on
Function keys show but no waveform source ↗
- Front-end-to-motherboard comms: replace the cable, then the front-end board, then the motherboard
Keyboard dead / LCD frozen source ↗
- Remove AC; hold On/Off 10 s; wait 10 s and restart
- Or remove the battery fuse for 10 s
RS-232 or modem transfer fails source ↗
- Match the baud rate to the receiving system; otherwise the cable or motherboard
- Modem: verify the phone number and init string
Print quality / writer issues source ↗
- Writer test: Alt+Shift+RHY on AC
- Check the 200 mm line ±2 mm (motor), missing dots (print head) and the cue sensor trace
Speaker volume or date/time not retained source ↗
- F6 More → 3 Date & Time → Alt+Shift+C → Volume 0–8; replace the speaker, then the motherboard
- If date/time isn't kept after saving with F5, replace the motherboard
🧰 Common checks for ECG (2)
Leads off / bad waveform source ↗
- Swap patient cable, then harness, then ECG board
Power/battery source ↗
- Check AC fuses; fewer than 10 reports per charge = replace battery
📞 Still stuck? Call Baxter
Contact Baxter US
- 📞 Tech support
- 1-800-535-6663, option 2
- 🛠️ Service
- 1-800-445-3720 (Hillrom beds/stretchers/lifts parts, service, tech support)
- [email protected]
- 🌐 Portal
- Support portal ↗
- 🔩 Parts
- Order parts ↗
Welch Allyn monitoring/cardiology: 800-535-6663 opt 2. Hillrom beds service email [email protected], parts [email protected]. Sigma Spectrum pumps: 1-800-843-7867 (catalog.baxter.com/baxterUS/en/contactus). Fee-based tech support for beds/stretchers without contract since 2025-03-31.
Missing a manual, code, or fix for this model? Tell us.