Connex 360
Baxter · Vital Signs Monitors
Open your document
Open ↗
Assembly instructions - Welch Allyn Connex 360 Mobile Stand
Welch Allyn Connex 360 Vital Signs Monitor - Instructions for Use
Stuck? Start here
✅ Check the simple stuff first
- Make sure the patient is safe — swap in a backup unit if you need to.
- Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
- Check power, battery, cables, accessories, and settings before ordering parts.
- Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for Connex 360 (6)
Won't power up source ↗
- New battery: connect AC to wake it
- Power supply output should be 15 V ±0.45 V DC
- Charge 5 h; replace battery (battery PCBA kit 108962) / main board
Touchscreen unresponsive or blank source ↗
- Touch to wake from Sleep
- Calibrate; run Host Calibration Sequence in Service Tool
- Replace LCD/touch assembly (kit 108872), then main board
No nurse-call alarm signalling source ↗
- Check alarm limits and that audio isn't silenced
- Host Calibration Sequence; main board
Won't join Wi-Fi source ↗
- Check network settings and range
- Run Blackwell radio functional check; update software
- Replace radio PCBA (kit 108873) / antenna (1097242)
Ethernet won't connect source ↗
- Verify network port configured
- Run Ethernet functional check; update software; main board
NIBP not working source ↗
- Leak-check cuff and hose; check single-lumen switch PCBA (kit 108963)
- Replace NIBP module (kit 108837)
🔢 Connex 360 error codes (19)
| Code | Meaning | Next step |
|---|---|---|
| 050002 | NIBP not functional (measurement unavailable) | Internal fault: service if persists source ↗ |
| 050003 | Unable to determine NIBP: motion or wrong settings | Check patient mode/settings, limit movement source ↗ |
| 050004 | Excessive artifact | Check connections, restrict movement source ↗ |
| 050005 | Low inflation | Check inflation settings/patient mode source ↗ |
| 050006 | Kinked tubing or transducer calibration failure | Check hose for kinks; service source ↗ |
| 050007 | Deflation dumped too soon | Check settings/patient mode source ↗ |
| 050008 | Not enough steps | Check connections, restrict movement source ↗ |
| 050009 | Invalid patient info for mode | Check patient mode source ↗ |
| 05000A | Re-inflation too late | Check connections, restrict movement source ↗ |
| 05000B | Too many re-inflate attempts | Check inflation settings source ↗ |
| 05000C | Couldn't bleed below safe venous return pressure | Check tubing for kinks and connections source ↗ |
| 05000D | Air leak in cuff or tubing | Check cuff and hose connections source ↗ |
| 05000F | Auto-zero failure | Pressure unstable: check connections; service source ↗ |
| 050105–05020A | NIBP module communication/message errors | Internal: service source ↗ |
| 050503 | Factory EEPROM checksum error | Service source ↗ |
| 050504 | User EEPROM checksum error | Calibrate NIBP; service if persists source ↗ |
| 050505 | A/D converter POST failure | Service source ↗ |
| 050509 | Calibration signature zero | Calibrate NIBP module source ↗ |
| 050512 | Invalid NIBP algorithm | Verify algorithm setting source ↗ |
🧰 Common checks for Vital Signs Monitors (2)
Won't power on source ↗
- Confirm AC cord/outlet and battery charge
- Check internal power/battery harnesses
- Measure power supply output against spec (15 V DC on Welch Allyn Connex and Philips VS3)
- Charge battery fully; replace if still empty
Touchscreen issues source ↗
- Check whether touch is disabled/locked
- Calibrate touchscreen
- Check touchscreen cables before replacing parts
📞 Still stuck? Call Baxter
Contact Baxter US
- 📞 Tech support
- 1-800-535-6663, option 2
- 🛠️ Service
- 1-800-445-3720 (Hillrom beds/stretchers/lifts parts, service, tech support)
- [email protected]
- 🌐 Portal
- Support portal ↗
- 🔩 Parts
- Order parts ↗
Welch Allyn monitoring/cardiology: 800-535-6663 opt 2. Hillrom beds service email [email protected], parts [email protected]. Sigma Spectrum pumps: 1-800-843-7867 (catalog.baxter.com/baxterUS/en/contactus). Fee-based tech support for beds/stretchers without contract since 2025-03-31.
Missing a manual, code, or fix for this model? Tell us.