Connex 360

Baxter · Vital Signs Monitors

Open your document

Welch Allyn Connex 360 Vital Signs Monitor - Service Manual

Service ManualOEM officialEN
Open ↗

Assembly instructions - Welch Allyn Connex 360 Mobile Stand

Other DocumentOEM officialEN
Open ↗

Welch Allyn Connex 360 Vital Signs Monitor - Instructions for Use

Operator's ManualOEM officialEN
Open ↗

Stuck? Start here

✅ Check the simple stuff first
  1. Make sure the patient is safe — swap in a backup unit if you need to.
  2. Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
  3. Check power, battery, cables, accessories, and settings before ordering parts.
  4. Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for Connex 360 (6)

Won't power up source ↗

  • New battery: connect AC to wake it
  • Power supply output should be 15 V ±0.45 V DC
  • Charge 5 h; replace battery (battery PCBA kit 108962) / main board

Touchscreen unresponsive or blank source ↗

  • Touch to wake from Sleep
  • Calibrate; run Host Calibration Sequence in Service Tool
  • Replace LCD/touch assembly (kit 108872), then main board

No nurse-call alarm signalling source ↗

  • Check alarm limits and that audio isn't silenced
  • Host Calibration Sequence; main board

Won't join Wi-Fi source ↗

  • Check network settings and range
  • Run Blackwell radio functional check; update software
  • Replace radio PCBA (kit 108873) / antenna (1097242)

Ethernet won't connect source ↗

  • Verify network port configured
  • Run Ethernet functional check; update software; main board

NIBP not working source ↗

  • Leak-check cuff and hose; check single-lumen switch PCBA (kit 108963)
  • Replace NIBP module (kit 108837)
🔢 Connex 360 error codes (19)
CodeMeaningNext step
050002NIBP not functional (measurement unavailable)Internal fault: service if persists source ↗
050003Unable to determine NIBP: motion or wrong settingsCheck patient mode/settings, limit movement source ↗
050004Excessive artifactCheck connections, restrict movement source ↗
050005Low inflationCheck inflation settings/patient mode source ↗
050006Kinked tubing or transducer calibration failureCheck hose for kinks; service source ↗
050007Deflation dumped too soonCheck settings/patient mode source ↗
050008Not enough stepsCheck connections, restrict movement source ↗
050009Invalid patient info for modeCheck patient mode source ↗
05000ARe-inflation too lateCheck connections, restrict movement source ↗
05000BToo many re-inflate attemptsCheck inflation settings source ↗
05000CCouldn't bleed below safe venous return pressureCheck tubing for kinks and connections source ↗
05000DAir leak in cuff or tubingCheck cuff and hose connections source ↗
05000FAuto-zero failurePressure unstable: check connections; service source ↗
050105–05020ANIBP module communication/message errorsInternal: service source ↗
050503Factory EEPROM checksum errorService source ↗
050504User EEPROM checksum errorCalibrate NIBP; service if persists source ↗
050505A/D converter POST failureService source ↗
050509Calibration signature zeroCalibrate NIBP module source ↗
050512Invalid NIBP algorithmVerify algorithm setting source ↗
🧰 Common checks for Vital Signs Monitors (2)

Won't power on source ↗

  • Confirm AC cord/outlet and battery charge
  • Check internal power/battery harnesses
  • Measure power supply output against spec (15 V DC on Welch Allyn Connex and Philips VS3)
  • Charge battery fully; replace if still empty

Touchscreen issues source ↗

  • Check whether touch is disabled/locked
  • Calibrate touchscreen
  • Check touchscreen cables before replacing parts
📞 Still stuck? Call Baxter

Contact Baxter US

📞 Tech support
1-800-535-6663, option 2
🛠️ Service
1-800-445-3720 (Hillrom beds/stretchers/lifts parts, service, tech support)
✉️ Email
[email protected]
🌐 Portal
Support portal ↗
🔩 Parts
Order parts ↗

Welch Allyn monitoring/cardiology: 800-535-6663 opt 2. Hillrom beds service email [email protected], parts [email protected]. Sigma Spectrum pumps: 1-800-843-7867 (catalog.baxter.com/baxterUS/en/contactus). Fee-based tech support for beds/stretchers without contract since 2025-03-31.

Source ↗

Missing a manual, code, or fix for this model? Tell us.