Connex VSM 6000

Baxter · Vital Signs Monitors

Open your document

Welch Allyn Connex Vital Signs Monitor 6000 Series Service Manual

Service ManualOEM officialP/N 80026025EN
Open ↗

Welch Allyn Connex Vital Signs Monitor 6000 Series Service Manual (SW 2.0X-2.4X)

Service ManualOEM officialP/N 80024005EN
Open ↗

Stuck? Start here

✅ Check the simple stuff first
  1. Make sure the patient is safe — swap in a backup unit if you need to.
  2. Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
  3. Check power, battery, cables, accessories, and settings before ordering parts.
  4. Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for Connex VSM 6000 (5)

Won't power on source ↗

  • Check power-flex cable at J6 on main board
  • Check AC harness IEC connector to power supply, and supply J2 to main board J30
  • Check battery harness (battery connector board J2 to main J29)
  • Power supply output should be 15 V ±0.45 V DC
  • Charge battery 5 h; replace if still empty; then main board

Battery won't charge / short run time source ↗

  • Charge 5 h, replace if icon stays empty
  • Inspect battery connector board for shorts/broken connector
  • Check cycle count in service tool; replace if over 300

Noisy fan source ↗

  • Blow dust out with canned air
  • Replace fan if out of balance

CO2: 'Filter line disconnected' source ↗

  • Connect/firmly seat the sampling line
  • Replace defective sampling line; clean input connector
  • If detection switch faulty, replace module

CO2: 'Occlusion in gas line' (high priority) source ↗

  • Replace the sampling line
  • If purge can't clear internal blockage, replace module
🧰 Common checks for Vital Signs Monitors (2)

Won't power on source ↗

  • Confirm AC cord/outlet and battery charge
  • Check internal power/battery harnesses
  • Measure power supply output against spec (15 V DC on Welch Allyn Connex and Philips VS3)
  • Charge battery fully; replace if still empty

Touchscreen issues source ↗

  • Check whether touch is disabled/locked
  • Calibrate touchscreen
  • Check touchscreen cables before replacing parts
📞 Still stuck? Call Baxter

Contact Baxter US

📞 Tech support
1-800-535-6663, option 2
🛠️ Service
1-800-445-3720 (Hillrom beds/stretchers/lifts parts, service, tech support)
✉️ Email
[email protected]
🌐 Portal
Support portal ↗
🔩 Parts
Order parts ↗

Welch Allyn monitoring/cardiology: 800-535-6663 opt 2. Hillrom beds service email [email protected], parts [email protected]. Sigma Spectrum pumps: 1-800-843-7867 (catalog.baxter.com/baxterUS/en/contactus). Fee-based tech support for beds/stretchers without contract since 2025-03-31.

Source ↗

Missing a manual, code, or fix for this model? Tell us.