Spot LXI Power Supply Bracket Service Installation Kit, Assembly Instructions

Other DocumentOEM officialEN
Open ↗

Spot LXI Vital Signs Monitor Battery Replacement, Installation Guide, Multi-Language

Other DocumentOEM officialEN
Open ↗

Stuck? Start here

✅ Check the simple stuff first
  1. Make sure the patient is safe — swap in a backup unit if you need to.
  2. Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
  3. Check power, battery, cables, accessories, and settings before ordering parts.
  4. Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for Spot LXI (6)

NIBP C04 air leak / won't reach pressure source ↗

  • Leak-check cuff, hose and connectors
  • Look for kinked or obstructed tubing (also causes C02 auto-zero)

NIBP C05 motion/noise errors source ↗

  • Re-position correct-size cuff; limit patient movement

E11 residual pressure source ↗

  • Verify power supply voltage within spec; retry

E20 internal source ↗

  • A/D failure: call Welch Allyn Technical Support

Temperature: probe errors (event 32-34) source ↗

  • Thermistor pulled/heater broken: retry predict then replace probe
  • Probe not calibrated: replace probe

C13 battery failure source ↗

  • Use wall transformer; replace battery
🔢 Spot LXI error codes (12)
CodeMeaningNext step
C01User abort— source ↗
C02Auto-zero failure / couldn't bleed below 9 mmHgCheck tubing for kinks/obstruction, connections; limit movement source ↗
C03Inflation too quickCheck tubing and connections source ↗
C04Major air leak / couldn't reach target pressureExcess inflation time: check for air leaks source ↗
C05Excessive arm movement / noisy steps / cycle too longCheck cuff placement; restrict movement source ↗
C06Reading outside measurement range / reinflation neededCheck connections; restrict movement; retry source ↗
C07Auto cycle skippedDevice will power down source ↗
C12Ambient temperature out of rangeCheck probe-well temp and ambient spec source ↗
C13Battery failureRun on wall transformer; replace battery source ↗
E10OverpressureCheck patient source ↗
E11Malfunction / residual pressure (>10 mmHg or >15 mmHg for >3 min)Retry; verify supply voltage within spec source ↗
E20Missed A/D sample or A/D timeoutInternal malfunction: call Technical Support source ↗
🧰 Common checks for Vital Signs Monitors (2)

Won't power on source ↗

  • Confirm AC cord/outlet and battery charge
  • Check internal power/battery harnesses
  • Measure power supply output against spec (15 V DC on Welch Allyn Connex and Philips VS3)
  • Charge battery fully; replace if still empty

Touchscreen issues source ↗

  • Check whether touch is disabled/locked
  • Calibrate touchscreen
  • Check touchscreen cables before replacing parts
📞 Still stuck? Call Baxter

Contact Baxter US

📞 Tech support
1-800-535-6663, option 2
🛠️ Service
1-800-445-3720 (Hillrom beds/stretchers/lifts parts, service, tech support)
✉️ Email
[email protected]
🌐 Portal
Support portal ↗
🔩 Parts
Order parts ↗

Welch Allyn monitoring/cardiology: 800-535-6663 opt 2. Hillrom beds service email [email protected], parts [email protected]. Sigma Spectrum pumps: 1-800-843-7867 (catalog.baxter.com/baxterUS/en/contactus). Fee-based tech support for beds/stretchers without contract since 2025-03-31.

Source ↗