✅ Check the simple stuff first
  1. Make sure the patient is safe — swap in a backup unit if you need to.
  2. Reproduce the complaint yourself and write down exactly what you see (screen, alarm, code).
  3. Check power, battery, cables, accessories, and settings before ordering parts.
  4. Then open the service manual above and follow its troubleshooting section.
🛠️ Common faults for SURETEMP PLUS (7)

Battery indications source ↗

  • Blinking frame + 'b' arrow = battery out of range; single blinking segment = low
  • Replace batteries

'A' with arrow source ↗

  • Ambient outside 10–40 °C: move to normal room temp

Probe or instrument with an X source ↗

  • Probe X = probe fault, usually replace probe
  • Instrument X = internal fault; read code in Biotech mode Error Log

Reading out of tolerance (calibration check) source ↗

  • Verify with the 9600 Plus Calibration Tester (01802-110) or the SureTemp Calibration Key (06138-000)
  • Field repairs are limited to calibration check, probe/well, battery and unit replacement

Need software version or error log (Biotech mode) source ↗

  • 692: place it in the wall holder; hold Recall while pulling the probe, then release
  • Recall steps through the categories (device/algorithm version, ID, error log); Mode or C/F changes values; returning the probe exits; times out after 5 min

Battery drains during cleaning or storage source ↗

  • Unplug the latching probe connector when removing the well; reconnect until it snaps in

Wrong probe or well, or probe damaged source ↗

  • Blue well for oral/axillary (02891-000), red for rectal (02891-100)
  • Probe kits: oral 02893-000 (4 ft) / -100 (9 ft); rectal 02892-000 / -100
  • Use only Welch Allyn covers (05031-750); biting the tip damages the probe
🔢 SURETEMP PLUS error codes (15)
CodeMeaningNext step
E0.1Excessive heater energyRetry; then replace probe source ↗
E0.2A/D out of rangeKeep ambient 10–40 °C; then replace probe source ↗
E0.4Probe over temperature (>43.3 °C)Let tip cool; then replace probe source ↗
E0.5No probe temperature riseRetry; then replace probe source ↗
E0.6Probe EEPROM read failureFresh batteries; then replace probe source ↗
E0.7Probe EEPROM un-initializedFresh batteries; then replace probe source ↗
E0.8EEPROM CRC failureFresh batteries; then contact Welch Allyn source ↗
E0.9Bad probe calibrationUse a calibrated probe source ↗
E4.0PTB measurement out of rangeRetry; then contact Welch Allyn source ↗
E4.1RCAL measurement out of rangeRetry; then contact Welch Allyn source ↗
E4.2Reference supply malfunctionContact Welch Allyn source ↗
E4.3–E4.7Device EEPROM save/handshake/response/transmit/init failureFresh batteries; then contact Welch Allyn source ↗
E4.8Device not calibratedReplace with calibrated device source ↗
E5.0Heater not workingRetry; then contact Welch Allyn source ↗
E5.2Heater fail-safe failureContact Welch Allyn source ↗
🧰 Common checks for Vital Signs Monitors (2)

Won't power on source ↗

  • Confirm AC cord/outlet and battery charge
  • Check internal power/battery harnesses
  • Measure power supply output against spec (15 V DC on Welch Allyn Connex and Philips VS3)
  • Charge battery fully; replace if still empty

Touchscreen issues source ↗

  • Check whether touch is disabled/locked
  • Calibrate touchscreen
  • Check touchscreen cables before replacing parts
📞 Still stuck? Call Baxter

Contact Baxter US

📞 Tech support
1-800-535-6663, option 2
🛠️ Service
1-800-445-3720 (Hillrom beds/stretchers/lifts parts, service, tech support)
✉️ Email
[email protected]
🌐 Portal
Support portal ↗
🔩 Parts
Order parts ↗

Welch Allyn monitoring/cardiology: 800-535-6663 opt 2. Hillrom beds service email [email protected], parts [email protected]. Sigma Spectrum pumps: 1-800-843-7867 (catalog.baxter.com/baxterUS/en/contactus). Fee-based tech support for beds/stretchers without contract since 2025-03-31.

Source ↗

Missing a manual, code, or fix for this model? Tell us.